ADS EOS Instructions For Use And Installation

Extraoral suction system

Advertisement

http://www.adsequip.com/
EOS Extraoral Suction System
Instructions For Use and Installation
P/N:8027313 Version:A Date of Issue:2020-05-06
Copyright©ADS Dental System. All Rights Reserved.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EOS and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ADS EOS

  • Page 1 EOS Extraoral Suction System Instructions For Use and Installation P/N:8027313 Version:A Date of Issue:2020-05-06 Copyright©ADS Dental System. All Rights Reserved.
  • Page 2 Please visit www.adsdental.com or www.adsequip.com for the latest product information. Once you have used the EOS system, it is understood that you have read this disclaimer and warning carefully, understood, recognized and accepted all terms and contents. It is your responsibility for the proper use of the system and agree to these terms and any regulations, policies and guidelines established by ADS.
  • Page 3: Table Of Contents

    EOS Extraoral Suction System, Instructions For Use and Installation Contents 1 Product Introduction…………………………………………………………………………………………………1 1.1 Intended Use of the Product…………………………………………………………………………………1 1.2 Product Standard Lifetime……………………………………………………………………………………1 Safety Precautions……………………………………………………………………………………………………1 3  Product Structure……………………………………………………………………………………………………4 4  Product Specifications………………………………………………………………………………………………5 4.1 Product Dimensions ……………………………………………………………………………………………6 4.2 Product Package Size and Weight …………………………………………………………………………8 5.
  • Page 4: 1 Product Introduction

    EOS Extraoral Suction System, Instructions For Use and Installation 1 Product Introduction The instructions contained within this manual should be thoroughly read and understood before operating the equipment. After the installation is completed, keep this manual in a safe place for future reference.
  • Page 5 EOS Extraoral Suction System, Instructions For Use and Installation CAUTION ADS will not be responsible when equipment damage or failure is caused by the below issues. •The system is not installed, modified, or maintained by ADS designated operator. •System damage or failure, is caused by products purchased from companies other than ADS authorized dealers.
  • Page 6 ④ Never expose the system to chemicals or place the it near a chemical storage area. ⑤ Clean and disinfect the system after every dental procedure. 7)Should a problem occur, please contact an ADS authorized dealers technical support team. Do not disassemble or attempt to repair.
  • Page 7: 3  Product Structure

    EOS Extraoral Suction System, Instructions For Use and Installation 3 Product Structure 8026695 Case Panel 8026722 Handle 8026722 Fine Filter A121945 Suction Mouth Piece Hood 8026635 Motor 8026746 Suction arm of the third joint 8026623 UV-C Light A121944 The third joint...
  • Page 8: 4  Product Specifications

    EOS Extraoral Suction System, Instructions For Use and Installation 4 Product Specifications Model EOS Extraoral Suction System Electric Voltage AC110V 60Hz 12-20A current Power 1160W Fuse wire F25A 110V Suction Flow 105CFM 23KPa (10 Different Levels) Power Average Efficiency (EM) for 0.4MM particles (%),90<EM<95 Fine Filter Minimum efficiency* for 0.4MM particles (%),55...
  • Page 9: Product Dimensions

    EOS Extraoral Suction System, Instructions For Use and Installation 4.1 Product Dimensions Unit:inch...
  • Page 10 EOS Extraoral Suction System, Instructions For Use and Installation...
  • Page 11: Product Package Size And Weight

    EOS Extraoral Suction System, Instructions For Use and Installation 4.2 Product Package Size and Weight Suction arm Case Packing Size 27.2”x10.2”x9” 14.6”x14.2”x39.4” Net Weight 3.2lbs 94.2lbs Gross Weight 5lbs 98.6lbs 5 Product Installation 1.After unpacking, open the two door panels and remove the four fixed parts.
  • Page 12 EOS Extraoral Suction System, Instructions For Use and Installation 2. Affix the bottom of center post to the base cabinet with the screws provided. Place tube through the hose clamp and connect it to the center post. Then tighten the clamp.
  • Page 13: Product Use

    EOS Extraoral Suction System, Instructions For Use and Installation 3. Insert the mouthpiece tube into the mounting hole of the third joint. Make sure the thumb screw is aligned with the center pillar groove. Tighten the screw. 6 Product Use 6.1 Warnings During Use...
  • Page 14 EOS Extraoral Suction System, Instructions For Use and Installation Handle Place the product on Move by handles level ground...
  • Page 15: Preparation

    EOS Extraoral Suction System, Instructions For Use and Installation 6.2 Preparation 110V 60Hz (20A) Ø Connect the 3 Prong Power Cord to the wall receptacle for 110V 20A power supply.
  • Page 16: Turn On The Power

    EOS Extraoral Suction System, Instructions For Use and Installation 6.3 Turn On the Power • Remove anything that might be inadvertently sucked in to the system. • Make sure the suction tubes are installed properly. • Press the “Power” button to turn the equipment on.
  • Page 17: Operation Principle

    EOS Extraoral Suction System, Instructions For Use and Installation 6.5 Operation Principle The EOS system collects aerosols, droplets, dust and pathogens produced during routine dental procedures through a suction mouthpiece hood, aerosols and droplets are collected into a well sealed metal box which is including a F8 fine filter, motor, UVC light and a H14 HEPA filter.
  • Page 18: Parts Cleaning And Replacement

    EOS Extraoral Suction System, Instructions For Use and Installation 7 Parts Cleaning and Replacement • •When installing or removing the suction mouth piece hood, hold it by the connecting end, instead of by the far end. • •To clean the equipment surface, use a disinfectant wipe and dry with a soft cloth.
  • Page 19: Instructions For Replacing The Suction Arm

    EOS Extraoral Suction System, Instructions For Use and Installation 7.2 Suction Arms [Daily Clean] •Use a spray disinfectant or wipes to clean the systems surface. •To internally clean the system, turn it on to its lowest suction level, spray VIROS OPTIM1 hydrogen peroxide disinfectant into the suction tubes and let the system run for 3 minutes.
  • Page 20: Instructions For Replacing The Noise Filter Stick

    EOS Extraoral Suction System, Instructions For Use and Installation 7.3 Noise Filter Stick Replacement (replace time :12 months) •Cut off the main power and remove power cord. •Keep your hands dry. •Wear gloves. [Caution] Replaced the noise filter stick and dispose of as medical waste.
  • Page 21: Instructions For Replacing The Hepa Filter

    EOS Extraoral Suction System, Instructions For Use and Installation 7.5 HEPA Filter Replacement Percentage of remaining lifetime (replace time :12 months) The lifetime of the HEPA filter is 12 months. When the equipment is powered on, the percentage of filter’s remaining lifetime will be shown on the panel.
  • Page 22: Instructions For Replacing The Uv Light

    EOS Extraoral Suction System, Instructions For Use and Installation UV –C Light Replacement •Turn off the systems power and remove powercord from wall. •Keep your hands dry. •Wear gloves. ① Remove door panel. ② Unlock the dust-proof cabinet drawer and replace the cover plate.
  • Page 23: Instructions For Replacing The Fuse Wire

    EOS Extraoral Suction System, Instructions For Use and Installation 7.7 Fuse Replacement •Turn off power switch, remove power cord. •Keep your hands dry. ① Remove power cord from wall. ② Open the cradle cover using a Philips screwdriver as instructed on fuse cradle.
  • Page 24: Transportation And Storage Conditions

    ADS Customer Services. 11 Warranty and Customer Services n The EOS Extraoral suction system comes with 2-year parts warranty from the date of purchase. Should you need to service your system, please contact the local dealer authorized by ADS.
  • Page 25 EOS Extraoral Suction System, Instructions For Use and Installation 8026695 Case Panel 8026722 Handle 8026722 Fine Filter A121945 Suction Mouth Piece Hood 8026635 Motor 8026746 Suction arm of the third joint 8026623 UV-C Light A121944 The third joint 8026563 HEPA filter...
  • Page 26: Electro Magnetic Compatibility

    For the Extraoral Suction System, use cables that are authorized by ADS only. Attempt to use cable or other components from unauthorized source could lead to electromagnetic irregularities.
  • Page 27 EOS Extraoral Suction System, Instructions For Use and Installation Statement of Manufacturer— Electromagnetic Immunity Extraoral Suction System is designed for electromagnetic environment described below. Be sure to apply. Electromagnetic Electromagnetic IEC6061 Test Level Test Level Conformance Immunity Test Environment The floor should be of wood, concrete, or tile.
  • Page 28 EOS Extraoral Suction System, Instructions For Use and Installation Statement of Manufacturer— Electromagnetic Immunity Extraoral Suction System is designed for electromagnetic environment described below. Be sure to apply. Electromagnetic IEC6061 Test Level Test Level Conformance Electromagnetic Environment Immunity Test The isolation distance between...
  • Page 29 EOS Extraoral Suction System, Instructions For Use and Installation Recommended isolation distance between portable and mobile RF communication devices and Extraoral Suction System Extraoral Suction System are intended for use in electromagnetic environments where RF radiation disturbances are controlled. Depending on the maximum rated output power of the communication device, the purchaser or user can prevent electromagnetic interference by maintaining the minimum isolation distance between portable and mobile RF communication devices (transmitters) and Extraoral Suction System as recommended below.
  • Page 30: Symbols Descriptions

    EOS Extraoral Suction System, Instructions For Use and Installation 14. Symbols Descriptions Symbols Descriptions Handle with Care Keep Dry More Information Serial Number Grounding Connection This Way Up Caution Type B Machine Power Switch Start/Stop Gear Up Gear Down UV Light...
  • Page 31 EOS Extraoral Suction System, Instructions For Use and Installation...
  • Page 32 ADS Dental System Inc. Technical Support:800 488 9708 Tel: (626) 620 0456 E-mail: ads@adsequip.com www.adsdental.com www.adsequip.com Add: 1590 S MILLIKEN AVE,UNIT A, ONTARIO,CALIFORNIA,91761,USA P/N:8027313 Version:A Date of Issue:2020-05-06 Copyright©ADS Dental System. All Rights Reserved.