Download Print this page

General Specifications - Fluke III Series Instruction Sheet

Hide thumbs

Advertisement

General Specifications

Maximum Voltage Between
any Terminal and Earth
Ground
Display
Response Time of Digital
Display
Operating Temperature
Storage Temperature
Temperature Coefficient
Electromagnetic Compatibility
in RF field of 3 V/m on all
functions except MΩ
Relative Humidity
except 32 MΩ Range:
32 MΩ Range only:
Altitude
Battery Type
Battery Life
Continuity Beeper
Shock, Vibration
Size (H x W x L)
Weight
Safety
EMC Regulations
Certifications/Listings
600 V
Digital: 3,200 counts, updates 2.5/sec
Analog: 31 segments, updates 25/sec
V ac < 2 s
V dc < 1 s
Ω < 1 s to 320 kΩ, < 2s to 3.2 MΩ,
< 10 s to 32 MΩ
0°C to 50°C
-40°C to 60°C
0.1 x (specified accuracy) / °C
(<18°C or >28°C)
Total accuracy = Specified accuracy
plus +2.0 % of range for all functions
except:
320.0 mV range: total accuracy =
specified accuracy + 1.0 % of range
320.0Ω range: total accuracy =
specified accuracy + 8 % of range
0 % to 90 % (0 °C to 35 °C)
0 % to 70 % (35 °C to 50 °C)
0 % to 80 % (0 °C to 35 °C)
0 % to 70 % (35 °C to 50 °C)
Operating: 2000 meters
Storage: 12,000 meters
9 V, NEDA 1604 or 6F22 or 006P,
or NEDA 1604A or 6LR61
2000 hrs typical with alkaline
1600 hrs typical with carbon zinc
4096 Hz
per MIL-T-PRF 28800F Class III,
Sinusoidal, Non Operating
2.8 cm x 7.5 cm x 16.6 cm
(1.12 in x 2.95 in x 6.55 in)
340 g (12.0 oz)
600 V CAT II per ANSI/ISA S82.01-
1994, EN 61010-1: 1993, CSA C22.2
No 1010.1-92, UL 3111-1.
EN 61326-1: 1997

Hide quick links:

Advertisement

loading

  Related Manuals for Fluke III Series

This manual is also suitable for:

73 iii