Operation and maintenance

6.1
Overview
This section contains information on commissioning and operating Location
Anchors.
Field Communicator and AMS Device Manager instructions are provided for
convenience.
6.2
Instructions and procedure safety
Instructions and procedures in this section may require special precautions
to ensure the safety of the personnel performing the operations.
Information that potentially raises safety issues is indicated by a warning
symbol (
operation preceded by this symbol.
6.3
Battery replacement
WARNING
The Emerson Location Anchor shall be used only with the battery
(00G45-9000-0001) supplied by Rosemount. This battery has been officially
tested with the Location Anchor as required by the I.S. standards during the
assessment of the Emerson Location Anchor. The battery is not replaceable
in a hazardous location. Dispose of battery in accordance with local and
national requirements.
Procedure
1. To replace the battery, remove devices from hazardous area.
2. Remove enclosure cover.
3. Switch the Location Anchor to "OFF".
4. Loosen the screw holding the electronics assembly to the enclosure.
5. Release battery connection from electronics board.
6. Loosen the two screws on the battery holder and slide the battery
holder to the left.
Note
The screws holding down the electronics board do not need to be
removed, just loosened. Take care not to let the battery fall out of
enclosure.