32
and a rise time of 25 nsec. to "EXT. CHOPPED"
(BU2).
— With apeak-to-peak value of .9 V, the driver gene-
rator should operate and two time-base lines should
be written.
O. ALTERNATE
— Set "MODE" (SKS) to position "ALTERNATE".
— Set time base A to its free -running position and ad-
just the time co-efficient to ".1 sec./cm".
— Check that the driver generator is switched over by
the ALTERNATE pulse of time base A. In this case,
two lines should be written.
As distinct from position "CHOPPED" these two
time-base lines should be written continuously, but
alternately after each other.
— Check this also if the time co-efficient of time base A
has been adjusted to .OS µsec.
Remark
In case of breakdown, one can always apply to the
world-wide PHILIPS Service Organisation.
Whenever it is desired to send the apparatus to a
PHILIPS Service Centre for repairs, the following
points should be observed.
tie on a label, bearing full name and address of the
sender.
indicate as complete as possible the symptoms) of
the fault(s).
— carefully pack the apparatus in the original packing,
or, if no longer available, in a wooden crate.
send the apparatus to the address provided by your
local PHILIPS representative.