CALIBRATION PROCEDURE
NI 6738/6739
Analog Output Devices and Modules with NI-DAQmx
This document contains the verification and adjustment procedures for the National Instruments
PCIe-6738 and PXIe-6738/6739 analog output devices and modules. For more information
about calibration solutions, visit
Contents
Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Password ................................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 4
Calibration Procedure ............................................................................................................... 4
Initial Setup............................................................................................................................... 5
Self-Calibration ................................................................................................................ 5
Verification ............................................................................................................................... 7
Counter Verification ......................................................................................................... 11
Adjustment................................................................................................................................ 12
EEPROM Update ..................................................................................................................... 14
Reverification ........................................................................................................................... 14
Test Limits................................................................................................................................ 15
Software
Calibrating NI 6738/6739 devices and modules requires the installation of NI-DAQmx on the
calibration system. You must use NI-DAQmx 17.6 or later. Refer to the NI-DAQmx Readme to
ensure support for the device you want to calibrate. You can download NI-DAQmx from
ni.com/downloads
LabVIEW, LabWindows
NI-DAQmx, you need to only install support for the application software that you intend to use.
ni.com/calibration
. NI-DAQmx supports many programming languages, including
™
™
/CVI
, C/C++, C#, and Visual Basic .NET. When you install
.