5 Noraxon TeleMyo as Master and MVN Analyze as slave ..... 9 Start and Stop a recording ................9 5.1.1 Hardware connections ................9 5.1.2 Noraxon MyoResearch Software .............. 9 5.1.3 XSens Software ................... 10 6 Using the Link System ............. 11 www.xsens.com...
1 Synchronization Workflow The steps described below show how to make it possible for Xsens Awinda or Sync Station to send a signal (Awinda / Sync Station is Sync OUT and Noraxon is Sync IN) and how to receive the synchronization signal (Noraxon is Sync OUT, Xsens Awinda / Sync Station is Sync IN).
Sync OUT Sync OUT enables the Xsens system to send a trigger pulse via the Awinda or Sync Station to third party hardware. As with Sync IN, a combination of events are possible, based on a number of parameters.
Page 6
Figure 5 - Noraxon Telemoyo Mini-Receiver. The TeleMyo™ 2400R G2 Mini Receiver is an optional accessory item for the portable TeleMyo 2400T G2 Transmitter System. It is TTL compatible, facilitating sync inputs and outputs. For more detailed information, see the TeleoMyo user manual. www.xsens.com...
4.1.1 Hardware connections • In addition to the normal MTw hardware setup, connect the BNC connector from the TeleMyo DTS to Sync OUT 1 on the Awinda Station. • Connect the USB connection between the TeleMyo DTS and the PC.
Page 8
Figure 7 - Noraxon system ready to begin recording. When this screen is reached, click record, on the bottom left hand side of the screen on the Noraxon software. Figure 8 - Following “Record” command, system calibrates and waits for trigger input. www.xsens.com...
4.1.3 Xsens Software Initialize the synchronization setting of the Xsens software using the following settings. 1. In the Configuration window, in the Sync tab you should select: Strat recording (out) Stop Recording (out) • Out 1 • Out 1 •...
USB cables are required to connect each system to the PC. Note that if the signal received by the Xsens Sync or Awinda Station is 5V, a 5V- 3.3V SMD level translator is advised to prevent damage to the Station.
The rest of the software setup is the same as described above. However, instead of clickingRecord in Xsens software, one should click the hardware trigger supplied by Noraxon, to generate a manual trigger to both systems. Figure 9 - Configuration of TeleMyo mini-receiver in Noraxon MyoResearch software.
6 Using the Link System To use the XSens Link system instead of the Awinda system you need to have an Awinda Station. Then you should activate the Awinda station in the “Motion Capture Configuration window” by turn on the bottom on the sync station configuration tab. In this tab you can find all the details described above.