Maintenance

To prevent possible electrical shock, fire, or
personal injury:
Remove the input signals before you
clean the Product.
Do not operate the Product with covers
removed or the case open. Hazardous
voltage exposure is possible.
Use only specified replacement parts.
Have an approved technician repair the
Product.

General Maintenance

Periodically wipe the case with a damp cloth and mild
detergent. Do not use abrasives or solvents.
Dirt or moisture in the terminals can affect readings and
can falsely activate the Input Alert feature. Clean the
terminals as follows:
1.
Turn the Meter off and remove all test leads.
2.
Shake out any dirt that may be in the terminals.
3.
Soak a clean swab with mild detergent and water.
Work the swab around in each terminal. Dry each
terminal using canned air to force the water and
detergent out of the terminals.
info@Fluke-Direct.ca
XWWarning
Fluke-Direct

Fuse Test

As shown in Figure 10, with the Meter in the
function, insert a test lead into the
probe tip on the other end of the test lead against the
metal of the current input jack. If "" appears in the
display, the probe tip has been inserted too far into the
amps input jack. Back the lead out a bit until the message
disappears and either OL or a resistance reading appears
in the display. The resistance value should be as shown
in Figure 10. If the tests give readings other than those
shown, have the Meter serviced.
XWWarning
To prevent possible electrical shock, fire, or
personal injury:
Replace a blown fuse with exact
replacement only for continued
protection against arc flash.
Use only specified replacement fuses.
.ca
Digital Multimeter
Maintenance
jack and place the
33
1.800.561.8187