If you are using the A terminal, set the rotary switch
3.
to mA/A. If you are using the mA/μA terminal, set the
rotary switch to
for currents above 6000 μA.
(6 mA), or
4.
To measure dc current, press .
5.
Break the circuit path to be tested. Touch the black
probe to the more negative side of the break; touch
the red probe to the more positive side of the break.
Reversing the leads will produce a negative reading,
but will not damage the Meter.
6.
Turn on power to the circuit; then read the display.
Be sure to note the unit given at the right side of the
display (μA, mA, or A).
7.
Turn off power to the circuit and discharge all high-
voltage capacitors. Remove the Meter and restore
the circuit to normal operation.
24
info@Fluke-Direct.ca
for currents below 6000 μA
Fluke-Direct
The following are some tips for measuring current:
If the current reading is 0 and you are sure the Meter
is set up correctly, test the Meter's fuses as
described under "Testing the Fuses".
A current meter drops a small voltage across itself,
which might affect circuit operation. You can
calculate this burden voltage using the values listed
in the specifications in the Input Characteristics table.