USER’S GUIDE Congratulations! ! On buying the device, be sure to check the delivery set, presence and correctness of the Quality Warranty Card filling, the acceptance certificate and product selling marks. ! Please, thoroughly read the user’s guide before using the device. Keep the User’s guide for future use.
Page 3
ENDOEST-MOTOR 6.2.1. Accumulators charging………………………………….……………….. 6.2.2. Function of accumulators charge control…………..…………………….. 6.2.3. Power saving function……...…………………………………………….. 7. Using…………………………………………………………………... 7.1. Power switching on…………………………………………………………. 7.2. Calibration of the micromotor system …….…………………………..…….. 7.3. Selection of the Ni-Ti system, working program and file …….……………... 7.4. The device settings………………….………………………………………..
USER’S GUIDE 1. GENERAL INFORMATION 1.1. The device brief summary «EndoEst-Motor” is a compact device combining endodontic motor with microprocessor control for working with rotating Ni-Ti instruments and apexlocator for localization of root canal apical constriction (apex). The device operates in three modes: only in endomotor, only in apexlocator or mixed mode when the apexlocation function controls operation of endomotor .
ENDOEST-MOTOR 1.2. Field of application: The device is designed for use in dentistry (endodontics) and can be used only in hospitals by medical specialists licensed to practice dentistry. The manufacturer is not responsible for the device misuse. 1.3. Contraindications Do not use the device on patients with pacemakers.
Page 6
USER’S GUIDE to find the cause of malfunction basing on the information represented in part 13 of this user’s guide. ! This endomotor has function of the file limit torque restriction what lessens probability of break during its use. In spite of it, files may break down because of wrongly set limit torque value in the program or because of worn out files.
Page 7
ENDOEST-MOTOR ! The apexlocator malfunction is possible when using it in the area of strong electromagnetic interference (EMC). Don’t use the device close to electromagnetic waves emitting equipment. Malfunctions may take place close to the equipment marked with symbol. ! Do not use this device in conjunction with other equipment, or as part of other equipment.
USER’S GUIDE 2. DELIVERY SET The device delivery set is represented in the table 2 and fig. 1 Table 2. The device delivery set № on Position name Number of pieces in the set Variant Variant fig.1 “НР” “ММ” Control unit Micromotor “DC-НР”...
ENDOEST-MOTOR 3. ACCESSORIES Additional accessories for the «EndoEst-Motor” are represented in the table Table 3 Outward Number of Position name Pcs. in appearance order the pack GE99.069.000 Micromotor “DC-HP” Micromotor for working with rotating Ni-Ti instruments in the set with the head GЕ99.124.000...
Page 10
Input voltage: (220±10%) V, ~50Hz; Output voltage: 9V; 500 mA. Plug– USBB GE99.110.000 Accumulator unit is an additional Li-Po accumulator unit for the «EndoEst-Motor” (2*3,7V) GE99.112.000 Stand “Stand-Master” is a stand for the micromotor “DC- HP” and “DC-MM” GE99.113.000 Stand «Stand-Satellite»...
ENDOEST-MOTOR Extension of table 3 Outward Number of Position name Pcs. in appearance order the pack GЕ255.000.000 Endodontic stand «EstTray» is a stand for the micromotor “DC- HP” and “DC-MM” with clean-stand GE99.061.000 Foam rubber sponges for the clean-stand ∅...
Page 12
USER’S GUIDE Measurement range in the apexlocator mode - from 3,1 to – 0,4; Adjustment range of the apical shelf point position in the apexlocator mode - from 0,1 to 0,5 (by increment 0,1); Measurements accuracy in the apical area (from 0,5 to the apex) with probability 87% - ±...
ENDOEST-MOTOR 5. THE DEVICE DESCRIPTION. OPERATING CONTROLS AND INDICATIONS 5.1. Control unit Outward appearance of the «EndoEst-Motor” control unit is represented in the figure 2. Figure 2. Control unit 1а Where: 1 - case: а - front panel, b - back panel;, c - cover of the battery compartment (accumulator unit);...
USER’S GUIDE 7 - socket miniUSB for the “USB-miniUSB” cable connecting (10, fig.1) (for connecting the device control unit with computer in order to update the device firmware version); 8 - socket “MICROMOTOR” for connecting the micromotor “DC-HP” (2, fig.1) or “DC-MM” (3, fig.1); 9 - socket “PEDAL”...
ENDOEST-MOTOR Figure 4. Indication on the display in the apexlocator mode. а) example 1 b) example 2 where: 1 - preset localized point in the canal: АL - “Apex” or (0.1- 0.5) Sh - ‘Shelf” (see p. 7.4.2 D - setting “Locatable point”);...
Page 16
USER’S GUIDE Table 4. The control buttons description Name Functional designation fig. «POWER» - The device switching on/off - Control of the power batteries current charge level «SET» - Entering the mode of program general settings (menu “Settings”)/Reset of the endomotor current position mode «POWER»...
ENDOEST-MOTOR 5.1.3. Program library for the Ni-Ti file systems The device has a large “standard” programs library for working with the Ni- Ti file systems of the leading manufacturers with preset rotation speed and limit torque value for each file.
USER’S GUIDE Extension of table 5 Name of the Name of the working programs Ni-Ti system Main Additional Standard / Apical files/ Retreatment EndoFlare НЕRO 642 Easy canal / Average canal / Difficult canal Individual НЕRO Shaper Easy canal / Average canal / Difficult canal Individual НЕRO Apical Revo-S...
Page 19
ENDOEST-MOTOR Table 6. Sound indication Action Type of sound signal Power switch on/off Two-time signal Endomotor Micromotor start/stop One-time signal operation Current rotation limit torque Discontinuous signal of the file is 70% higher than the preset value Micromotor operation in the Signal of type “rattle”...
ENDOEST-MOTOR indication (see table 7) what simplifies the endomotor and apexlocator operation tracing without looking at the device display. Table 7. Light indication on the micromotor Operation Action Lightning on mode micromotor Current rotation limit torque of the file is 70%...
USER’S GUIDE 6. MAKEREADY After the device transportation at temperature below +5С you should let it warm up for an hour at room temperature. 6.1. Components connecting 6.1.1. Wired components connecting А) Micromotor “DC-HP” / “DC-ММ” (see fig.8) - Place the micromotor cable plug Fig.
ENDOEST-MOTOR - Place the connectable cable component plug towards the socket “APEX/ CHARGE” (6, fig.2) so that the canted and right angles of the socket both sides coincide (canted angles from bottom, right angles from above). - Insert the plug into the socket until bumping Note: To disconnect the wired components take the insulating part of the plug and pull it with small effort.
USER’S GUIDE - Shift the file fixing lever on the head over to the open outside right position (position “open”); - Set the file in the clamping hole until bumping, slightly turning the file until it is compatible with the fixing mechanism;...
ENDOEST-MOTOR Attention! Be sure to sterilize the apexlocator electrodes before using it before each patient (see part 10 “Caring for the device. Information on sterilization and disinfection”). Note: 1.When working in the mixed mode (endomotor with apexlocator) it’s not necessary to connect the file clip with the cable. The micromotor with working file serves to be an active electrode itself.
USER’S GUIDE 6.2. Accumulators Two chargeable Li-Ion accumulators are the «EndoEst-Motor” device power source. Fully charge the device accumulators before the first use. Note: The device may operate from the accumulator or the electric power source (until the accumulators are being charged) only in the endomotor mode.
Page 27
ENDOEST-MOTOR А) Indication of accumulators current charge level Every time switching on the device (see p.7.1), you will see displaying information of the accumulators current charge level in percent (with accuracy to 5%) (see example fig.17). Note: When the accumulators are being charged, you will see the word “Charge”...
Page 28
USER’S GUIDE 7. USING 7.1. Power switching on - Press the button “POWER” to switch on the device. Two informational messages indicate on the display following one another when switching on the device: Message 1. Hello message with manufacturer’s name and number of the current firmware version (fig.
Page 29
ENDOEST-MOTOR Note: The calibration of micromotor system is of prime importance when working with small torques in the canal apical area. Function of calibration (see fig.19) activates automatically when connecting the device “DC- HP” micromotor (fig. 6) (or the “DC-MM” (fig.7)) since the device power is switched on.
Page 30
USER’S GUIDE - Press the button “START/STOP” (3, fig.5 or 3, fig.6) or the foot control pedal (8, fig.1) to start the micromotor. - Calibration of micromotor system takes several seconds (fig.20). Please, wait until the calibration graphic scale on the display is completely full. - After the calibration system successful performance, the message “Calibration.
Page 31
ENDOEST-MOTOR Press the button “ENTER” to confirm your selection. B) Working program selection (fig.21b) Select the working program in the additional menu of the endomotor mode (if available) (see table 5, column 2). Use the buttons «>» or «<» (for fast search of the necessary program press and hold down one of the buttons) to select the program.
Page 32
USER’S GUIDE Select the necessary file in the endomotor working program (see p.7.3C) and press the button “ENTER” to enter the mode of the file individual settings. Herewith, indication in the figure 22 (as exemplified on the “ProTaper Universal” system, file S1-Sx) appears on the display screen.
Page 33
ENDOEST-MOTOR Attention! When selecting the file rotation speed value, follow the recommendations of the Ni-Ti instruments manufacturer. Do not preset speed values higher than the maximally allowed by manufacturer. Adjustment specifics of the file rotation speed in the “Standard” programs Maximal rotation speed of the file in the “standard”...
Page 34
USER’S GUIDE Maximally allowed limit torque value of the file in the endomotor “Basic” program depends on the rotation speed preset for it (see p.7.4.1.A). Approximate maximal limit torque values at a preset speed are indicated in the table 9. Table 9 Preset speed, rpm <=300...
Page 35
ENDOEST-MOTOR Press the button “SET” in any endomotor mode window to enter the general settings mode Use the buttons «>» or «<» to navigate through the menu “Settings” (fig.23) (for fast search of the necessary setting press and hold down one of the buttons).
Page 36
USER’S GUIDE On default the function “Autotwist” is activated in the program general settings. Use the buttons «>» or «<» to change the current setting Press the button “ENTER” to save the settings. Herewith the message “Setting is saved” indicates on the display.
Page 37
ENDOEST-MOTOR 2. Function “Autostop” Fig.27 When attaining the apex/ apical shelf, the micromotor stops rotating automatically - see fig.27 “Stop”. 3. Function “Continue” When attaining the apex/ apical s h e l f , t h e m i c r o m o t o r...
Page 38
USER’S GUIDE а) 2. “Shelf” When working with the apexlocator, distance measuring from the working file top to the apical shelf point on the preset distance (from 0,1 to 0,5) from the apex - see fig.29b. On default the value “Apex 0.0” is activated in the program general settings.
Page 39
ENDOEST-MOTOR F) “Language” (fig.31) This program version has two user interface languages - russian (“Русский”) and english. On default the program has English language interface activated in the general settings mode. Use the buttons «>» or «<» to change the current setting ...
Page 40
USER’S GUIDE Variant 2 (only at presence of the micromotor “DC-HP” and/or foot control pedal): Press and hold down the button “START/STOP” on the micromotor “DC- HP” case or the foot control pedal to start the micromotor Release the button or pedal to stop the micromotor 7.6.
Page 41
ENDOEST-MOTOR Herewith, indication (examples are represented in the figure 4) appear on the display. The apexlocator switches off automatically when opening the electrical circuit (the lip clip removing from patient’s lip, file extraction from the canal etc.). Herewith, the indication on the display recovers its state.
Page 42
USER’S GUIDE 8. STEP-BY-STEP INSTRUCTION ON THE DEVICE USING IN ALL THE OPERATING MODES 8.1. The device using in the endomotor mode 1. Connect the “DC-HP” micromotor (or “DC-MM” micromotor) to the device control unit (for more details see p.6.1.1A, page 22). 2.
Page 43
ENDOEST-MOTOR 10. Use the button “START/STOP” or the foot control pedal to start the micromotor after inserting the file into the root canal (for more details see p.7.5, pages 39-40). 11. Follow the file current limit torque in the graphic scale indicated on the display during the micromotor operation.
Page 44
USER’S GUIDE 3. Press the button “POWER” to switch the device on (for more details see 7.1, page.28) 4. Check the electrical circuit continuity of the apexlocator (for more details see p.7.6.1, page 40) 5. Insert the working file into the patient’s root canal (at the distance not more than 3,1 to the localized point) and activate the apexlocator mode in the following way: - place the lip clip on patient’s lip;...
Page 45
ENDOEST-MOTOR 1. See p.8.1 (1-8) and p.8.2 (1) 2. Connect already cleaned and sterilized lip clip to the cable socket. Herewith, the second cable socket may stay not activated (for more details see p.6.1.4, pages 24-25 and part 10 ) 3.
Page 46
USER’S GUIDE 9. FUNDAMENTAL OPERATING INSTRUCTIONS OF THE APEXLOCATOR For more accurate apexlocator measurement results, please, follow these rules: 1. Carefully isolate the studied tooth and working file from saliva. We recommend to use cofferdam, cotton swabs, saliva ejectors; 2. Work only in rubber gloves; 3.
Page 47
ENDOEST-MOTOR If blood, saliva, or a chemical solution overflows from the opening of the root canal and contacts gums, this will result in electrical leakage and an accurate measurement can not be obtained. Wait for bleeding to stop completely. Clean the inside and opening of the canal thoroughly to get rid of all blood, saliva and chemical solutions and then make a measurement.
Page 48
USER’S GUIDE 6. Crown or metal prosthesis touching gingival tissue (fig.41) Accurate measurement cannot be obtained if the file touches a metal prosthesis that is touching gingival tissue. In this case, widen the opening at the top of the crown so that the file will not touch the metal prosthesis before taking a measurement.
Page 49
ENDOEST-MOTOR 10. CARING FOR THE DEVICE. INFORMATION ON STERILIZATION AND DISINFECTION 10.1. Micromotor head lubricating The micromotor head (2, fig.6,7; 1,fig.46) should lubricated before each sterilization time. For the lubrication of the head it is recommend to use a standard aerosol can with a lubricant (3, fig.46) and a special...
Page 50
USER’S GUIDE ultrasound baths “UltraEst”, “UltraEst-FSM” or “UltraEst-M” manufactured by the “Geosoft-Dent”. The belongings sterilization should be carried out directly before the first device use and after each patient to avoid cross infection. All the belongings pointed should have autoclave sterilization (water saturated vapor under manometer pressure).
Page 51
ENDOEST-MOTOR 10.4. Replacing the foam rubber sponge in the clean-stand To replace the foam rubber sponge in the clean-stand you should: - take the clean-stand (4, fig.16) out of the “Stand-Satellite” (1, fig.16); - separate the clean-stand upper part (1, fig.47) from the lower part (2, fig.47) and extract the old...
Page 52
USER’S GUIDE 7. Set a new accumulator unit at the place of the used one а) 8. Assemble the device in reverse sequence. 9. Switch the device on and make sure it operates (see p.7.1) . 10. Connect the mains charger (see p.6.1.1C) and fully charge the accumulator unit (see p.6.2.1).
Page 53
ENDOEST-MOTOR Note: The replaceable seal ring is not in the device delivery set and may be acquired apart extra cost. (see part “Accessories”). 12. FIRMWARE UPDATING Attention! Internet access is needed for the device firmware updating. Return to the previous firmware version is impossible after updating it.
Page 54
USER’S GUIDE 13. TROUBLESHOOTING Table 10. General malfunctions Malfunction Cause Action The accumulators are Charge the accumulators (see The device doesn’t switch on discharged p.6.2.1) There is no accumulator Check presence and correctness unit or it’s not connected of the accumulator unit in the battery connection (see p.
Page 55
ENDOEST-MOTOR Extension of table 10 Malfunction Cause Action Sound signal volume See setting “Volume” (p. Problems with sound indication level is incorrectly set 7.4.2E) There is a bad contact Check connection of the The micromotor does not rotate...
Page 56
USER’S GUIDE Extension of table 10 Malfunction Cause Action The apexlocator is See p.7.6.2 The apexlocator doesn’t switch on incorrectly connected The apexlocator Check the apexlocator electrical electrical circuit circuit continuity (see p.7.6.1) and continuity is broken remove the trouble (see table 11) ...
Page 57
ENDOEST-MOTOR Table 11. Troubleshooting in the case of the apexlocator electrical circuit continuity break. № Action Indication Indication didn’t appeared appear (see fig. 32) apexlocator mixed mode mode Check if the apexlocator cable connection ОК See p. 3 See p.2 with the control unit is correct and reliable (see p.6.1.1С, page22) and the lip clip and...
Page 58
USER’S GUIDE 14. STORAGE CONDITIONS, TRANSPORTATION AND USE The device should be stored in heated and ventilated locations at temperature from +5°С to +40 °С with relative air humidity of 80% (at +25° С) in the authentic package of the manufacturer. ...
Page 60
USER’S GUIDE 17. WARRANTY 1.The manufacturer guarantees the device successful operation according to the technical specification requirements, technical conditions 9452-009- 56755207-2005 when the user follows all the regulations, service and storage conditions. 2. Warranty period of the device use is 12 months since the day of sale, but not more than 18 months since the day of manufacturing (except p.3).
Page 61
ENDOEST-MOTOR 18. SYMBOL DESCRIPTIONS Symbol Description Warning: Address to supporting documentation Type of protection against electric-shock hazard. Device of the II class Protection level from electrical shock: The device of type В Continuous current Do not throw away the device into system of daily rubbish...
Page 62
APPENDIX Electromagnetic Emissions and Immunity Table 1 The device “EndoEst-Motor” is intended for use in the electromagnetic environment specified below. The customer ore the user of the device should assure that it is used in such an environment. Emission test...
Page 63
ENDOEST-MOTOR Table 2 The device “EndoEst-Motor” is intended for use in the electromagnetic environment specified below. The customer ore the user of the device should assure that it is used in such an environment. Immunity test Test level Compliance Electromagnetic...
Page 64
USER’S GUIDE Continuation of Table 2 Immunity test Test level Compliance Electromagnetic EN 60601-1-2 Level environment - guidance Voltage dips, < 5 % UT < 5 % UT Mains power short (> 95 % dip in (> 95 % dip in quality should be interruptions UT) for 0.5 and 1...
Page 65
ENDOEST-MOTOR Table 3 The device “EndoEst-Motor” is intended for use in the electromagnetic environment specified below. The customer ore the user of the device should assure that it is used in such an environment. Immunity Test level Complianc Electromagnetic environment -...
Page 66
Recommended working clearances between portable and mobile RF communication devices and the device “EndoEst-Motor” The device “EndoEst-Motor” is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by...
Page 67
Quality Warranty Card № 1 Date of repair Malfunction Service Master Quality Warranty Card № 2 Date of repair Malfunction Service Master Quality Warranty Card № 3 Date of repair Malfunction Service Master...
Page 68
“ENDOEST-MOTOR” Serial number: Control unit Stamp Micromotor Date of manufacture Date of sale Seller “ENDOEST-MOTOR” Serial number: Control unit Stamp Micromotor Date of manufacture Date of sale Seller “ENDOEST-MOTOR” Serial number: Control unit Stamp Micromotor Date of manufacture Date of sale...
Page 69
ENDOEST-MOTOR Certificate of acceptance Set variant НР ММ (mark the appropriate) Control unit Micromotor Date of manufacture Version Controller The stamp of the manufacturer Sales mark Date of sale Seller Stamp of trading organization Sales mark is obligatory! The warranty is invalid without sales mark.
Need help?
Do you have a question about the ENDOEST-MOTOR and is the answer not in the manual?
Questions and answers