Introduction The CENTRY™ 117 Microcentrifuge is a compact, touchscreen, benchtop microcentrifuge designed for separation of various research samples. The motor is brushless and requires no routine maintenance. The microcentrifuge is supplied with a CoMBi-Rotor™ for 24 microtubes (1.5–2.0 mL) and two PCR-strips for micro volume samples.
Any claims for damage must be filed within 48 hours with the carrier. NOTE The accessories and instruction manual supplied with the CENTRY 117 Microcentrifuge should be kept with the near its place of installation. Installation Site Requirements The CENTRY 117 Microcentrifuge should be installed on a rigid, even surface such as a stable laboratory bench, countertop, etc.
Do not attempt to open the lid of the CENTRY 117 WARNING Microcentrifuge until the rotor has come to a complete stop.
Page 8
2. Remove the plastic plug, located on the left side of the unit. 3. Pull the wire (attached to the plug) to open the lid lock manually. Do not attempt to override the lid lock mechanism. Doing so is dangerous and could damage the CENTRY 117 WARNING Microcentrifuge. Speed Selection...
Page 9
Start/Stop button. The microcentrifuge will continue to run as long as the button is held down. Stop a Run To stop the CENTRY 117 Microcentrifuge prior to the expiration of set time, press the Start/Stop button. CENTRY™ 117 MiCRoCENTRifugE | uSER’S guiDE...
Maintenance Rotor Maintenance The rotor should be cleaned regularly. Thorough cleaning must be performed when spinning samples containing phenol or phenol chloroform. Periodically inspect the rotor for dents, dings, scratches, discoloration, and cracks. if any damage to the rotor is found, discontinue use of the rotor immediately and replace.
Page 11
Repairs performed by unauthorized personnel may void the warranty. Always keep the CENTRY 117 Microcentrifuge housing, rotor chamber, rotor, and rotor accessories clean. All parts should be wiped down periodically with a soft cloth. For more thorough cleaning, use a neutral cleaning agent (pH between 6 and 8) applied with a soft cloth.
Page 12
Replace Fuses Check the fuse when recommended (refer to Troubleshooting on page 13). 1. Disconnect the power cord from the power inlet. 2. Locate the fuse holder in the power inlet on the rear panel. 3. open the fuse holder drawer by inserting a small screwdriver under the tab and prying it open.
Check that power is being supplied to the outlet Check that the power cord is plugged into No power supply CENTRY 117 both the wall outlet Microcentrifuge will and the back of the not start centrifuge Check that power cord...
Page 14
No power from PC Call for service board Lid lock is jammed Call for service Lid lock will not release See CENTRY 117 Microcentrifuge is not Microcentrifuge will receiving power not start on page Lid not closed Close lid correctly...
This warranty is valid only if the instrument is used in the manner described in this guide and for the purpose for which it is designed. gilson is not responsible for consequential damages resulting from the misuse or bad cleaning or decontamination of this instrument.