1. Introduction
Thank you for choosing Champion S-33 centrifuge from Ample Scientific.
Champion S-33 features:
Durable and sophisticated design
Temperature controlled motor to extend centrifuge life
expectancy
Temperature controlled inner chamber through vortex
ventilation system providing stable centrifuging environment
Stainless steel inner chamber
Ultra smooth acceleration curve
2. Product Structure
Rotor
Power switch is
located in the
back
Speed Control
and Display
Figure 1 - Product External Structure
Pg | 1
Lid
Centrifuge Chamber
[On] button to start
centrifugation
Manual unlock opening
[Off] button to stop
centrifugation and to
open lid
Timer control
and display
3. Installation and Setup
1. Inspect the centrifuge for any damages. Contact distributor for
questions regarding damages.
2. Place centrifuge on steady and leveled surface. The area should be
well ventilated, and the centrifuge should be kept at least 300mm
away from surrounding objects in all directions. S-33 is for indoor
use only.
3. Open the centrifuge lid:
a. Connect centrifuge to power source and turn on power
switch from the back of the centrifuge. When panel lights
up, press [Off] for the lid to automatically unlock. Listen for
the 'click' sound before opening the lid. If the centrifuge
does not unlock, proceed to [b.].
b. Using the provided screw driver, insert into small opening
on the right side of the centrifuge just below the lid;
maneuver until a 'click' is heard and the lid is unlocked.
4. Remove packaging material in the inner chamber
5. Place buckets into rotor. Place circular rubber inserts into rotor
buckets. Remove any debris or pieces of left over packaging
material from inner chamber.
6. Close lid and make sure it is locked. Select time for 1 minute. Press
[On] to start operation.
7. Make sure the whirring sound produced is smooth. If there is
excessive vibration or abnormal sounds, stop the centrifuge
immediately. Make sure rotor is correctly installed and there is no
debris in centrifuge. If vibration and sound persist, please contact
your distributor immediately.
8. Your centrifuge is now ready to be used.
Pg | 2

Advertisement

loading
Need help?

Need help?

Do you have a question about the Champion S-33 and is the answer not in the manual?

Questions and answers

Paul
March 19, 2025

What is the size of the tubes? How many mls. per tube?

1 comments:
Mr. Anderson
May 14, 2025

The specific size and capacity in milliliters (ml) of the tubes for the Ample Scientific Champion S-33 centrifuge is not provided in the given context. However, it mentions that the rotor buckets are designed to hold most commonly sized tubes, vacutainers, and SSTs, and that some tubes may be too large for the rotor buckets and should not be used.

This answer is automatically generated

Summary of Contents for Ample Scientific Champion S-33

  • Page 1 1. Introduction 3. Installation and Setup Thank you for choosing Champion S-33 centrifuge from Ample Scientific. 1. Inspect the centrifuge for any damages. Contact distributor for Champion S-33 features: questions regarding damages. 2. Place centrifuge on steady and leveled surface. The area should be ...
  • Page 2 4. Centrifuge Operation Note: the lid will remain unlock for 3 seconds only. Afterwards, please press [off] to unlock and open the lid. 1. Turn on centrifuge using power switch located on the back of centrifuge. The display panels should light up, accompanied by a long ‘beep’.
  • Page 3 6. The centrifuge should be kept and stored in the following condition for optimal performance: a. Ambient temperature of 0-30˚C b. Relative humidity of ≤ 30% c. Atmosphere clear of electroconductive dusts, explosive gas, and corrosives 6. Prohibited Usage of Centrifuge 1.
  • Page 4 Champion S-33 Centrifuge User Guide Ample Scientific L.L.C. www.amplescientific.com...