7 Calibrated Measuring Amplifier (C1250 -xx-XC-xS- Cxx)
The drives with the ending -Cxx are specially designed for measuring applications. They come with a factory
calibration certificate for the analog inputs on X4. The analog inputs on X4 provide a measuring error of less
than 1%.
These drives have at least hardware version 1 revision H4 (V1RH4) and therefore come as well with an UL
listing.
The firmware must be at least version 6.9.
It is the user's responsibility to allow a reasonable period for recalibration. We recommend a calibration interval
of 12 months.

8 Description of the connectors / Interfaces

8.1 PE
PE
Protective Earth
PE
8.2 X1
X1
Motor Supply
PWR
PGND
Connector has to
Motor Supply: 72 VDC nominal, 24...85 VDC
be ordered
Absolute max. Rating: 72 VDC +20%.
separately: see
External Circuit Breaker: 15 A / min. 100 VDC / C-Trip / 5 kA rms SCCR
chapter 15)
* for cURus: External Fuse: 16A slow blow / min. 100VDC
If motor supply voltage exceeds 90 VDC, the drive will go into error state.
Page 10 of 32
Use min. 4 mm
2
(AWG11)
Tightening torque: 2 Nm (18 lbin)
Use 60/75 °C copper conductors only
Conductor Cross-Section 2.5 mm
0185-1063-E_6V16_IG_Drives_C1250
2
(AWG14) max Length 3 m