COOK Medical Evolution Instructions For Use Manual page 7

Esophageal stent system partially covered/fully covered
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3
is not possible to recapture stent after passing point-of-no-return, indicated when red marker
on top of introducer has passed the point-of-no-return indicator on handle. (See Fig. 2)
7. To recapture stent, push direction button on side of delivery system to opposite side. (See
Fig. 3) Note: Hold thumb on button when squeezing trigger for first time to recapture.
8. Continue squeezing trigger as required to recapture stent by desired amount.
9. To resume deployment, push button to opposite side again and hold thumb on button for
first stroke while squeezing trigger. (See Fig. 4)
10. When stent point-of-no-return has been passed, pull safety wire out of delivery handle
near wire guide port. (See Fig. 5)
11. Continue deploying stent by squeezing trigger.
12. After deployment, fluoroscopically confirm full stent expansion. Once full expansion is
confirmed, introduction system can be safely removed. Note: You may introduce endoscope
and advance to the upper margin of the stent to endoscopically confirm position and patency
of the stent. Warning: Do not introduce the endoscope into the stent as displacement may
occur.
For partially covered stent:
To reposition stent directly after placement, grip the stainless steel/
lasso at
13.
white
proximal-end of stent with forceps and reposition stent to
area. (See Fig. 6)
desired
For fully covered stent:
e
14. To reposition stent directly after placement, grip either the proximal-end lasso or th
distal-end la
sso with forceps and reposition stent to desired area. (See Fig. 7)
Upon completion of procedure, dispose of device per institutional guidelines for
biohazardous medical waste.
7