Diagnostics by ESP
Diagnostics by ESP
Click this button to bring
up the
Diagnostic
menu.
To open nested options,
single-click
the +.
To change a setting,
double-click
the
setting and use your
cursor to scroll
through the options.
Diagnostics by Serial Command
Power-on/Resets Counts
Hours Since Reset
Warning Messages
Laser High/Laser Low
High Temperature
Low Temperature
Service Message
12-2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Diagnostics
<K406,?> returns: <K406,power-on,resets,power-on saves,custom
default saves> (read only)
<K407?> returns: <K407,hours,minutes>
<K400,warning message status,laser high status,laser low sta-
tus,novram/reset warning status>
<K411,laser high status,laser high message,laser low status,laser low
message>
<K402,high temperature status,high temperature message>
<K403,low temperature status,low temperature message>
<K409,status,service message,threshold,resolution>