DENTSPLY Ray-pex 5 User Manual

Apexlocator
Hide thumbs

Advertisement

Quick Links

Apexlocator
User Manual

Advertisement

loading

  Related Manuals for DENTSPLY Ray-pex 5

  Summary of Contents for DENTSPLY Ray-pex 5

  • Page 1 Apexlocator User Manual...
  • Page 2: Table Of Contents

    Table of contents Introduction………………………………………………………. Package Contents………………………………………………. Short Description ……………………………………………..Charging the Battery ………………………..………………..Operation ………………………………..………………………. Troubleshooting ………………………………………………… 12 Demo Mode and Virtual Apex ………………………………… 17 Maintenance and Sterilization ……………………………….. 19 Technical Specifications ……………………………….……… 20 Symbols …………………………………………………………. 20 10. Conformity ……………………………………………………..21 11.
  • Page 3: Introduction

    ® 5 – one of the most innovative apex You have made a good choice with Ray-pex locators on the market. Based on the proven technology of the 4 generation devices, ® this apex locator has been optimized. Ray-pex 5 presents additional features, e.g. the special display graphics and its user-friendly ergonomic design.
  • Page 4: Package Contents

    Package Contents Before set-up, please check the contents of the package ® ● 1 Ray-pex 5 apex locator ● 1 charger, 1 measurement cable ● 1 tester for operation check ● 1 accessory set including: 2 lip clips, 2 file clips, 1 touch probe ●...
  • Page 5 ► Display Icons Measurement status – as long as the measurement circuit is open, the lip clip icon blinks. If the icon blinks during the measurement process, stop measuring and check lip clip and file clip connection. Measurement cable connection – indicates proper connection of the measurement cable to the device.
  • Page 6 Function and settings in Fig. 4 and 5. Display On-Off Demo Mode Volume Measurement cable/charger cable jack Electronic unit Fig. 4 Battery compartment Measurement cable compartment Service Interface Fig. 5 Rev. 3 01/2012...
  • Page 7: Charging The Battery

    Charging the battery Prior to first use, the battery must be charged for 24 hours ! ® Ray-pex 5 is powered by a NiMH rechargeable battery. The battery status is shown during measurement and while recharging. If the battery is low, the last segment on the battery indicator blinks and the battery ®...
  • Page 8: Operation

    Operation Important advice To isolate the working area always place a rubber dam. Prior to measurement the access cavity must be dried with a cotton pellet in order to prevent leaking current. Select the size of the measuring file to slightly grip at the apex. Always wear gloves during the measuring procedure.
  • Page 9 ► General Information ® The device does not need to be calibrated. Ray-pex 5 works with nearly all types of metal instruments, e.g. stainless steel and NiTi files (no strongly oxide-coated NiTi files). The size of the file must correspond to the size of the canal, i.e. it must fit the apical constriction.
  • Page 10 Fig. 7 ● Preapical and apical region On the right side of the display, the segments of the APEX ZOOM fill-up while the file advances (Fig. 8). Fig. 8 The segments show the position of the file tip: ● 4 blue segments: warning zone close to the apical region ●...
  • Page 11 The movement of the file along the apical region is accompanied by audio signals which increase in frequency the deeper the file advances. As soon as its tip reaches the apical foramen a constant beep will sound. Note: As with all electronic length determination devices, the scale shown in the display does not represent mm or inches! Once the file tip has passed the apical foramen the warning dot will appear and short warning beeps will sound (Fig.
  • Page 12 ► Interruption of measurement During the measurement cycle the file clip may be disconnected from the file and reconnected at any time without interrupting the measurement (e.g. when the file is changed to a larger size or when a second canal should be measured). The device detects automatically that a new measurement is initiated and signals it with two short beeps.
  • Page 13: Troubleshooting

    Please first review the below checklist in order to exclude any user error or anatomic/other peculiarities before contacting your dealer. If the problem persists please contact either your local dealer or Dentsply Limited directly.
  • Page 14 Is the lip clip connected? Check the cable and plug connections: Are the cable and plugs Are they plugged-in connected correctly? correctly? Is the measurement cable Check visible damage to damaged, e.g. torn? cable or plug? The electric circuit is not complete, i.e.
  • Page 15 Short circuit due to excess Dry the access cavity with liquid (irrigation solution, a cotton pellet / air-blower. saliva, blood) in the pulp chamber? In case of excess blooding wait until it can be stopped. For isolation: Is there a direct contact of the - adequate preparation measuring file with the gingiva filling...
  • Page 16 In these cases, ® it is important to use the Ray-pex 5 recording as the more accurate measurement. In the case of a lateral canal curvature the x-ray image may show a shorter working length than with Ray-pex 5. Rev. 3 01/2012...
  • Page 17 Take the tester plug out of the cable jack when check-up is completed. Warning: Readings outside of the green section indicate that the device is not functioning properly. In this case contact either your local distributor for assistance or directly Dentsply Limited. ► Checking the measuring cable (ONLY AFTER checking the device) Rev.
  • Page 18: Demo Mode And Virtual Apex

    Warning: Readings outside of the green section indicate that the device is not functioning properly. In this case contact either your local distributor for assistance or directly Dentsply Limited. Demo mode and Virtual apex ► Demo mode The built-in Demo mode allows you to become familiar with the device and to demonstrate the operation to your patients.
  • Page 19 ● Press Mode button again to interrupt the simulation. ● Press Mode button again to resume the simulation. ► Virtual apex The additional function of the virtual apex can be activated in the demo mode. This function enables you to set an individual apical measurement point. ●...
  • Page 20: Maintenance And Sterilization

    Maintenance and Sterilization ® The Ray-pex 5 does not require any routine maintenance and does not contain user serviceable parts. Repairs must be provided by trained service personnel only. Cleaning the measurement cable: we recommend to wipe it off with a cloth slightly moistened with a mild disinfecting solution (like Mikrozid AF or Meliseptol rapid).
  • Page 21: Technical Specifications

    Technical Specifications ® The Ray-pex 5 belongs to the following category of medical devices: Internal battery power (3.6V NiMH rechargeable battery) Protection against electric shock: Type BF Do not use in proximity of flammable anesthetic mixtures containing air, oxygen or nitrogen oxide Continuous operation No protection against ingress of liquids...
  • Page 22: Conformity

    (See declaration of conformity in the appendix). Notified Body: TÜV Produkt Service GmbH Ridlerstrasse 65 80339 Munich, Germany 11. Manufacturer Distributor VDW GmbH Dentsply Limited Bayerwaldstr.15 Dentsply International 81737 Munich Building 1, Aviator Park, Station Road Germany Addlestone Surrey, KT15 2PG...
  • Page 23: Warranty

    13. Warranty The manufacturer warrants this product against defects in material or workmanship for a period of 12 months from date of purchase. Cables and battery are covered by a warranty period of 6 months. During the warranty period a product determined by VDW to be defective will be repaired or, at the option of the manufacturer, will be replaced at no charge.
  • Page 24 Rev. 3 01/2012...
  • Page 25 Dentsply Limited Dentsply International Building 1, Aviator Park, Station Road Addlestone Surrey, KT15 2PG Tel.: +44 1932 853422 Fax: +44 1932 858970 www.dentsply.com Manufacturer: VDW GmbH Bayerwaldstr.15 * D-81737 Munich * Germany Tel: +49-(0)89-627340, Fax: +49-(0)89-62734190 www.vdw-dental.com * info@vdw-dental.com 01-2012...