Keysight N5247-60103 Installation Note page 32

Extended power range and bias tees upgrade kit

Advertisement

Step 22.
Reinstall the Inner Cover
For instructions, click the Chapter 7 bookmark "Removing the Covers" in the PDF Service Guide
Step 23.
Reinstall the Outer Cover
For instructions, click the Chapter 7 bookmark "Removing the Covers" in the PDF Service Guide
Step 24.
Enable Option 419
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A keyboard must be connected to the network analyzer.
Option Enable Procedure
1. To start the option enable utility, press UTILITY
option enable dialog box will appear.
2. Click the arrow in the Select Desired Option box. A list of available options will appear.
3. In the Select Desired Option list, click 419 - Src/Rcvr Atten & Bias Ts 4-Port.
4. Using the keyboard, enter the license key in the box provided. The license key is printed on the the license
message you received from Keysight. Enter this key exactly as it is printed on the message.
5. Click Enable.
6. Click Yes in answer to the displayed question in the Restart Analyzer? box.
7. When the installation is complete, click Exit.
Option Verification Procedure
Once the analyzer has restarted and the Network Analyzer program is again running:
1. On the analyzer's Help menu, click About Network Analyzer.
2. Verify that "419" is listed after "Options:" in the display. Click OK.
If Option 419 has not been enabled, perform the
NOTE
option is still not enabled, contact Keysight Technologies. Refer to
from Keysight" on page
Installation Note N5247-90103
Service
System
, then
"Option Enable Procedure"
3.
1
.
1
.
Option Enable
, then
. An
again. If the
"Getting Assistance
31