Toshiba LF414 Instruction Manual page 27

Electromagnetic flowmeter detector
Hide thumbs Also See for LF414:
Table of Contents

Advertisement

(2) Pipe Orientation
The detector may be installed in horizontal, vertical or sloping pipe runs as shown in
Figure 4.5. However, except for horizontal installation, fluid should flow from lower to
upper directions. If no air bubble, vertical down flow application are acceptable under
pressured piping conditions. See Figure 4.5.
Detector
(a)
The electrodes should be positioned horizontally relative to the ground surface in any piping
installation. See Figure 4.6.
Detector
(b)
(c)
Figure 4.5 Detector Piping Orientation
A
A'
Figure 4.6 Installation position of the detector
- 26 -
Flow direction
(a) Horizontal pipe installation
(b) Vertical pipe installation
(c) Sloping pipe installation
Ground surface
Electrodes
Cross-section A - A'
Ground surface
6F8A0870

Advertisement

Table of Contents
loading

Table of Contents