Electromagnetic flowmeter detector
Hide thumbs

Advertisement

Table of Contents
6F8A0870
(2) Preventing an Empty Pipe Condition
Fix the relevant pipes installed on both sides of the detector by attaching fittings, etc. to support the pipe. By
supporting the pipes, not only the pipe vibration is reduced but also the damage to the pipes by the
electromagnetic flowmeter's weight and the fluid mass (see Figures 4.2 and 4.3).
Pipe support fittings
Pipe support fittings
Figure 4.2 Example of Pipe Fixing Procedure
Figure 4.3 Model Diagram of Unsupported Pipes
- 22 -