Setting the data format
X
0
XYZ: Continuous output mode
1 2
X
=
X.Y.Z.: New line output mode
1 2
X
=
1 2
Y
=
1 2
Z
=
: Data is signed zero-suppressed 7-digit data (Space when sign is positive)
Data
1
: Display status (N: normal, D: double display)
2
: Display mode (I: INC, A: ABS)
3
: Unit (MM: mm, INCH: inch)
Note
• The data format during printer mode is
The data is signed zero-suppressed 6-digit X-axis data (space when the sign is positive), and the 7th digit is not
output. (
: Unit (M:mm, I:inch))
4
• Use printer mode when performing statistical data processing using the P30 (discontinued) or P40 printer.
(Statistical processing can only be performed for the X-axis).
• To output only the X, Y, and Z-axis data, COMP mode should be used with the P40 printer. COMP mode cannot
be used with the P30 printer since the communication process takes time. Data reception will become impossible
partway, and data will be lost.
Communication speed setting
X
0
Operating Procedure
Select the X-axis.
Press the
and
keys to switch the data
0
output format mode.
3
1 2 =
Y
Data
Space
3
CR
LF
Data
3
CR
LF
Data
3
CR
LF
Data
4 1
R
Operating Procedure
Select the X-axis.
Press the
key.
0
The communication speed increases.
Press the
key.
The communication speed decreases.
3
1 2 =
Z
Data
Space
CR
LF
.
Space
Data
Display
3
CR
LF
Data
Display
21