Using Oxygen Tubes (Only For Navy) - Dräger HP NO Series Instructions For Use Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 20

Using oxygen tubes (only for navy)

Slowly open the valve for the compressed air supply.
Provide the T-piece (1) with two pieces of hose, each 5cm long, and in-
sert in the Dräger tube holder H
Disconnect the rubber cap on the connection tube and break off both
tips.
Break off both tips on the oxygen tube.
Connect the oxygen tube (2) to a piece of hose about 5cm long.
Observe the green marking and arrow on the oxygen tube.
Connect the entry side of the oxygen tube to the T-piece.
Connect the entry side of the connection tube (4) to the hose (3).
Connect the output side of the connection tube to the accuro gas
detection pump. Note the associated Instructions for Use.
Press the accuro gas detection pump (5) together as far as it will go
once, then loosen it and wait for the end of the stroke.
Close the valve for the compressed air supply, immediately remove the
oxygen tube from the holder, and analyse the results in accordance with
the Instructions for Use.
The oxygen tube heats up during the measurement. Do not use in an
explosive area, as this is an explosion hazard.
Place the rubber cap on the connection tube. The connection tube can
be used four times.
Visit: www.thesafetyequipmentstore.com or email besafe@thesafetyequipmentstore.com for sales and service.
O.
2
CAUTION
4
Oil
CO
CO H
O
2
2
3
2
1
5
27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aerotest hpAlphaNavyMultitestHp usaHp nf

Table of Contents