Troubleshooting

Problem
Erratic trace
Erratic display
Signal quality
indicator is red
continuously
Questionable
FHR
FHR not recorded
Light or no trace
End of paper
noted when pack
not finished
Error message is displayed
If you suspect the signal from the transducer
If you suspect the signal from the recorder or display
Chapter 4 - Monitoring FHR and FMP Using Ultrasound
Possible Causes
Fetal arrhythmia
Obese patient
Transducer wrongly positioned
Belt loose
Too much gel
Very active fetus
Maternal movement
Insufficient gel
Transducer wrongly positioned
FHR less than 50 bpm
Recording MHR by mistake
Recording spurious periodic
signals when the transducer is not
applied to the patient
FHR exceeds 300 bpm
FHR is less than 50 bpm or
between 240 - 300 bpm
Wrong paper or dirty printhead
Bad paper feed or wrong paper
Solutions
None
None
Reposition transducer until signal quality
indicator is green
Tighten belt
Remove excess
None
Relax the patient
Use recommended amount
Reposition transducer until signal quality
indicator is green
None
Reposition transducer
FHR is half-counted (for example 320 bpm
is recorded as 160 bpm)
None
Use recommended paper or clean printhead
Check paper feed and use recommended
paper
See Chapter 15, "Troubleshooting" for a
table of error messages, their causes and their
solutions.
Carry out the Parameter Test as described on
page 115.
Carry out the Quick Test as described on
page 114.
Troubleshooting
35