3.2.5.1. Communication Setup Screen
The following appears on this screen (Figure 3-10):
The number on the left (e.g. 8) is the leak detector channel number.
The string on the right (e.g. 11:3C:0C) is the lower three-byte portion of the leak detector
MAC address. The MAC information is used to identify potential host leak detectors, or, if
more than one leak detector is set to the same channel, selecting the leak detector via
MAC address.
Once the desired MAC channel is shown in the Choose LD screen, pressing ENTER
wireless remote to attempt to establish a session with that leak detector. The leak detector and
wireless remote MAC addresses are viewed in the MAC Address screen (see TBD Section 2.2.5.3
"MAC Address Verification" on page 2-16).
To choose from an existing list:
1. Press the
Enter and the Speaker Volume screen appears (Figure 3-11).
Enter
Enter
Enter
Wireless Remote for Agilent Helium Leak Detector

Figure 3-10. Choose LD Screen

,
, +, or - hard buttons to highlight the Home screen Tools icon

Figure 3-11. Speaker Volume Setup Screen

ENTER causes the
ENTER
ENTER
and press
3-14