Control Accuracy

Parameter
Pressure
Inhale Time
Exhale Time
Pause Time
Pre-Therapy Breath
Time
Pre-Therapy Breath
Pause
Frequency
Amplitude
Device performance and accuracy is specified at Temperature: 20-30° C; Humidity: 50% relative; Altitude: nominally 380
meters for typical patients.
Sound
The sound pressure of the device set at -40 cmH
dBA at 1 meter.
Disposal
Dispose of the device in accordance with local regulations.
Range
-70 to 70 cmH
O
2
0-5 seconds
0-5 seconds
0-5 seconds
0 – 5 seconds
0 – 5 seconds
1-20 Hz
1-10 cmH
O
2
2
± 5 cmH
O
2
± (10% of setting + 0.1 second)
± (10% of setting + 0.1 second)
± (10% of setting + 0.1 second)
± (10% of setting + 0.1 seconds)
± (10% of setting + 0.1 seconds)
± (10% of setting)
± 5 cmH
O
2
O/+40 cmH
O in the Pause phase is less than 62
2
Chapter 8 Technical Specifications
Accuracy