After Installing HDD Data Encryption & Mirroring Kit
Installing the System Software Using the SST
The system data stored on the HDD and used to control the host machine will be lost when
the machine is first started up after installing this product. It is important to install the system
software used to control the host machine so that the machine may start up properly after
installation of this product. Details follow.
1. Requirements
1) PC
Service support tool in the version that supports this host machine must be installed.
2) Cross Ethernet Cable
2. Preparing for the Installation of the System Software of Host machine
1) If both PC and the machine are on, turn them off.
2) Connect the PC and the machine using an Ethernet cable.
3) Turn on the PC.
4) Start up the machine in download mode (safe mode).
3. Selecting the System Software
1) Set the CD containing the latest system software in the PC on which the SST is used.
2) Start up the SST.
3) Click 'Register Firmware'.
4) Select the drive in which the System Software CD has been set, and click 'SEARCH'.
5) Click 'REGISTER'.
6) Click OK.
4. Downloading the System Software
1) Click 'CONNECT'.
2) From the list of machine series, select the appropriate model.
3) Select 'Single', and click start.
4) Execute HDD format.
5) After 5 sec from when the power of the host machine is turned OFF, restart the host
machine in download mode of safe mode.
6) When "download mode" is displayed on the control panel, click simple mode start.
7) Click start to execute download.
8) Follow the instruction on the screen and when download is complete, click OK.
9) Exit SST.
9
Installation > [TYPE-8] Option HDD (80GB) + Removable HDD Kit + HDD Mirroring Kit or HDD Data Encryption & Mirroring Kit > After Installing H
10) Check the versions of MN-CONT and LANG etc in service mode (COPIER > Display >
VERSION).
Checking the Security Version
1) Press the Counter key (123 key) [1] on the control panel.
2) Press the [Check Device Configuration] key appearing on the control panel.
3) Make sure that '2.00' is displayed in 'Canon MFP Security Chip' as version information of
the security chip.
When several Encryption Boards are installed, multiple version information is displayed.
CAUTION:
The user will be able to make sure that the encryption board fitted with a security chip
of the correct version with CC authentication is functioning normally by referring to the
version information (2.00) indicated for 'Canon MFP Security Chip'.
Checking the Security Mark
The user may check the security mark, appearing on the control panel when using the host
machine to make sure that an appropriate level of security is being maintained. The mark
appears when the machine is equipped with an Encryption Board and the board is operating
correctly. The Users Guide provides the following description in connection with the security
mark:
<Confirming the Security Mark>
When the HDD Data Encryption & Mirroring Kit is operating normally, a security mark (
displayed on the lower left corner of a panel screen.
CAUTION:
The user will be able to make sure that the encryption board fitted with a security chip
of the correct version with CC authentication is functioning normally by referring to the
version information (2.00) indicated for 'Canon MFP Security Chip'.