Installation

CAUTION
ATTENTION
1.
Prepare the Lambda 365 and Peltier Controller in a location that is compatible with the required
environmental conditions for the operation.
2.
Open the lid of the coolant inlet on top of the Peltier Controller.
Figure 7 Lid of coolant inlet
3.
Fill up the liquid coolant.
CAUTION
ATTENTION
Keep the amount of liquid coolant to the level where the indicator is located between the 'L'(low) and
NOTE:
'H'(high) mark in the scale on the left side of the Peltier Controller. When the coolant is running short for
the operation, LED of Low Water Level is changed red with alarm sounding.
4.
Connect the power cord and communication cable of the Lambda 365. DO NOT turn on the power
of the instrument!
Make sure the instrument is turned off while installing this accessory.
Assurez-vous que l'instrument est éteint lors de l'installation de cet accessoire.
Liquid coolant should be used as refrigerant. Do not use any water (tab water,
DI, etc.). If water is used as refrigerant, the Peltier Controller's components might
be corroded and its performance may be deteriorated.
Use any kind of liquid coolant of normal grade that is available. (e.g. liquid coolant
including Distilled Water 70 wt%, Propylene Glycol 27 wt%, Additives 3 wt%).
Le liquide de refroidissement doit être utilisé comme réfrigérant. N'utilisez jamais
d'eau (eau du robinet, eau déminéralisée, etc.) L'eau risque en effet de corroder
les composants du contrôleur Peltier et de réduire les performances de ce
dernier.
Utilisez un liquide de refroidissement du commerce de qualité normale (par
exemple, un liquide de refroidissement à base d'eau distillée à 70 % M/V, de
propylène glycol à 27 % M/V et d'additifs à 3 % M/V).