EN 61010-1, EN 61010-2-020, EN 60529, EN 61326-1 and EN ISO 12100� › Keep the operating instructions in a place where it can be accessed by everyone� A copy of the complete EU Declaration of conformity can be requested at sales@ika�com� › Ensure that only trained staff work with the device�...
This also applies to any extra energy produced in other ways, e�g� through light irradiation� › The IKA G-L centrifuge is suitable for separating of substances and mixtures of substances of various densities� › Always use rotor cover along with the rotor�...
Open the safety lid of the device� Rotor nut User guide Program number display Display the selected program number� Rotor IKA CR 12 x 2 ml Warranty card “counter/timer” display Display the counter/timer value� Technical information IKA G-L Rotors “timer” indicator Indicate the timer function is activated�...
: maximum centrifugation radius in cm Example, The maximum centrifugation radius of the rotor IKA G-L CR 12x2ml is 6 cm� At a speed of 15700 rpm, a maximum g-force of 16500 x g can be achieved�...
Page 9
Centrifuging Pulse function Setting max. (15700 rpm) / Preset speed for Pulse function: › Set the desired rpm with “rcf/rpm” (+) button(D) / “rcf/rpm” (-) button (C)� › Open the lid by pressing lid release button (K)� › Press and hold Pulse Button (B)� ›...
Page 10
Editing the program: Activating the program: �...
Error codes and troubleshooting /// Warning/Note codes /// Error codes Causes › The load in the rotor is not balance Any malfunctions during operation will be identified by an error message on the display� Effect › motor off Solutions › switch off the device Proceed as follows in such cases: ›...
› device type� device� › serial number, see type plate › item and designation of the spare part, see: www.ika.com, spare parts diagram and spare parts list� › software version (Briefly visible in the display when the device is switched on)�...
See more accessories on www.ika.com� Install the driver by running the setup file� Connect the IKA device through the USB data cable to the PC� The data communication runs through a virtual COM port� /// Command syntax and format: The following applies to the command set: ›...
Technical data Warranty In accordance with IKA warranty conditions, the warranty period is 24 months� For claims under General data the warranty please contact your local dealer� You may also send the machine direct to our factory, enclosing the delivery invoice and giving reasons for the claim� You will be liable for freight costs�...
Page 15
Phone: +81 6 6730 6781 Phone: +91 80 26253 900 Phone: +44 1865 986 162 eMail: info@ika�in eMail: info_japan@ika�ne�jp eMail: sales�england@ika�com Discover and order the fascinating products of IKA online: www.ika.com IKAworldwide IKAworldwide /// #lookattheblue @IKAworldwide Technical specifications may be changed without prior notice�...
Need help?
Do you have a question about the G-L and is the answer not in the manual?