Optimizing the Waveform

Optimizing the Waveform
In the
minimum and maximum scales for the current wave.
Non-Physiological Artifact Suppression
Some clinical procedures may affect blood pressure, for example, a flush procedure or a blood sample.
Your monitor may be configured to suppress these non-physiological artifacts for a specified duration
(Artifact Suppression is configured to 30, 60, or 90 seconds). During artifact suppression,
the monitor shows the INOP message <Pressure label> ARTIFACT, and a question mark is
shown beside the pressure numerics. Pressure alarms and the Pulse Non-Pulsatile INOP are
suppressed during the configured period. The CPP high alarm is not suppressed.
Choosing the Pressure Alarm Source
WARNING Make sure you set alarm limits for the correct label. The alarm limits you set are stored for that
particular label only. Changing the label may change the alarm limits.
You can monitor for alarm conditions in systolic, diastolic and mean pressure, either singly or in
parallel. Only one alarm is given at a time, in this order of priority: mean, systolic, diastolic.
In the
Menu option
Sys.
Dia.
Mean
Sys&Dia
Dia&Mean
Sys&Mean
Sys&Dia&Mean
Select and set the
WARNING If you are using an intra-aortic balloon pump connected to the M1006B #C01 module, do not
defibrillate unless the pump cable is disconnected.
menu, select
Setup <Press>
menu, select
Setup <Press>
Pressure value monitored
systolic
diastolic
mean
systolic and diastolic in parallel
diastolic and mean in parallel
systolic and mean in parallel
all three pressures in parallel
High Limit
to let the monitor select the best
Optimum Scale
and choose the source.
Alarms from
and
for the pressure(s) you have selected.
Low Limit
12 Monitoring Invasive Pressure
133