Removing the AC Power Connector

To remove the AC power connector:
Step
1
2
3
4
Separate the front panel from the rear case assembly. See "Removing the Front Panel" on
page 5-11.
+
On the VS2
with the recorder option, remove the recorder. See "Removing the Optional
+
Recorder (VS2
Only)" on page 5-9.
Remove the cable tie that secures the power input cables to the chassis.
Disconnect the power input cable from the power supply assembly.
Power input cable
Removing the AC Power Connector
Repairing the Monitor
+
SureSigns VS2
and VSi Service Guide
5-19