Download  Print this page

Navigation Wheel (vs2 Monitor Only) And Key Problems; Recorder Problems (vs2 + Monitor Only) - Philips SureSigns VS2+ Service Manual

Vital signs monitor.
Hide thumbs

Advertisement

Troubleshooting Tables
Navigation Wheel (VS2
Symptom
The monitor does not respond to
the navigation wheel (VS2
or front panel keys.
Pressing a key does not access the
expected function. For example,
pressing the Print key initiates an
NBP measurement.
A key is collapsed, or doesn't
click when pressed.
Recorder Problems (VS2
Symptom
The recorder paper is jamming.
The recorder does not print.
Troubleshooting
+
4-10
SureSigns VS2
and VSi Service Guide
+
Monitor Only) and Key Problems
Possible Cause
The main board has a faulty
+
only)
connection.
The monitor keypad or navigation
wheel is broken.
The main board malfunctioned.
The membrane switch assembly
malfunctioned.
The membrane switch assembly
malfunctioned.
The membrane switch assembly
malfunctioned.
+
Monitor Only)
Possible Cause
The paper is not installed correctly.
The wrong type of paper is installed.
The recorder has a mechanical
problem.
An optional recorder is not installed.
The recorder is out of paper.
The recorder door is open.
The system does not recognize the
recorder.
The recorder module malfunctioned.
The main board malfunctioned.
Action
Ensure that all connectors to the main board
are seated correctly.
Replace the front panel or navigation
wheel. See "Removing the LCD" on
page 5-28 and "Replacing the Main Board"
on page 5-22.
Replace the main board. See "Replacing the
Main Board" on page 5-22.
Replace the front panel assembly. See
"Removing the LCD" on page 5-28.
Replace the front panel assembly. See
"Removing the LCD" on page 5-28.
Replace the front panel assembly. See
"Removing the LCD" on page 5-28.
Action
Remove paper and reinstall correctly.
Install the recommended recorder paper.
Replace the recorder module. See
"Removing the Optional Recorder (VS2
Only)" on page 5-9.
Verify that a recorder is installed.
Install a new paper supply.
Open the recorder door, and then close it
again.
Check the recorder's physical connections.
See "Removing the Optional Recorder
+
(VS2
Only)" on page 5-9.
Replace the recorder module. See
"Removing the Optional Recorder (VS2
Only)" on page 5-9.
Replace the main board. See "Replacing the
Main Board" on page 5-22.
+
+

Advertisement

Table of Contents

This manual is also suitable for:

Suresigns vsi