Display Problems

Symptom
The power is on, but the monitor
screen is blank.
The monitor displays random or
distorted graphics with a white
background.
The monitor turns on and the
opening screen and/or color bar
starts, but then freezes.

Alarm Problems

Symptom
A flashing red/white box with the
message
AUDIO FAILED
on the screen.
Audible alarms do not sound.
Possible Cause
The monitor is in Standby mode.
The main board malfunctioned.
The LCD is faulty.
The LCD cable is not attached
correctly.
The LCD malfunctioned.
The main board malfunctioned.
The main board malfunctioned.
Possible Cause
The speaker malfunctioned.
appears
The connection to the speaker is
faulty.
The speaker is faulty.
The monitor is in Audio Pause
mode or Audio Off mode.
The connection to the speaker is
faulty.
The speaker is faulty.
The main board malfunctioned.
Troubleshooting Tables
Action
Press the On/Standby key to end
Standby mode.
Replace the main board. See "Replacing
the Main Board" on page 5-22.
Replace the LCD. See "Removing the
LCD" on page 5-28.
Open the monitor and reseat the LCD
signal cable. See "Removing the LCD"
on page 5-28.
Replace the LCD. See "Removing the
LCD" on page 5-28.
Replace the main board. See"Replacing
the Main Board" on page 5-22.
Replace the main board. See "Replacing
the Main Board" on page 5-22.
Action
Clear the error message by performing the
speaker test or clearing the error log. See
"Testing the Speaker" on page 4-31 or
"Clearing Patient Data" on page 4-37.
Ensure that the connection to the speaker
is correct.
Clear the error message by performing the
speaker test or clearing the error log. See
"Testing the Speaker" on page 4-31 or
"Clearing Patient Data" on page 4-37.
Replace the speaker. See "Replacing the
Speaker" on page 5-15.
Clear the error message by performing the
speaker test. See "Testing the Speaker" on
page 4-31.
Confirm that audio alarms are enabled.
Ensure that the connection to the speaker
is correct.
Replace the speaker. See "Replacing the
Speaker" on page 5-15.
Replace the main board. See "Replacing
the Main Board" on page 5-22.
Troubleshooting
+
SureSigns VS2
and VSi Service Guide