Replacing The Fuse; Caring For Chemcassettes - Honeywell SPM Technical Handbook

Single point monitor
Table of Contents

Advertisement

14. Before securing cover to body, verify SPM will go
into Monitor Mode by powering-up the unit.
Note:
If SPM goes into Monitor Mode, proceed with
next step. If not, power down SPM, check all
cable connections and try again. If problem
persists, call Honeywell Analytics Service.
15. Power down SPM.
16. Secure cover to body (see
17. Power up unit and verify system response as
outlined in
Section

3.5 Replacing the Fuse

SPM operation is protected with a fuse located in a fuse
holder cap (33) at the lower left front panel. The fuse size
is 5 x 20 mm slo-blo. The fuse rating may vary based
upon installed SPM options. For replacement, refer to the
fuse designation that is listed on the fuse label placed
next to the fuse holder.
Warning
Disconnect the external power source before
checking or replacing the fuse. Use only the fuse
rated for your instrument. Refer to the front
panel fuse label for the proper fuse for your
instrument.
To replace the fuse, turn the fuse holder cap (33)
counterclockwise and pull out. The fuse will remain in
the cap.
Remove the blown fuse from the cap and replace with a
new fuse of the proper rating for your instrument. Insert
fuse into fuse holder and rotate cap clockwise.
SPM Technical Handbook
Section
2.5).
1.8.
SPM Single Point Monitor

3.6 Caring for Chemcassettes

The SPM uses an SP (standard play-2 week), or
EP (extended play-30 day) or XP (longer life-up to
three months) Chemcassette
Chemcassette
part numbers. Under normal conditions,
®
Chemcassettes
have a shelf life of approximately three
®
months for standard and 1 year for XP. They should
be stored in a cool atmosphere and kept out of direct
sunlight.
Most Chemcassette
optimum sensitivity when stored at room temperature
(16°C to 24°C; 60°F to 75°F). A few, such as Aliphatic
Amines/Ammonia, Ammonia, Chlorine III/Oxidizers,
Diisocyanates, Hydrogen Cyanide, Hydrogen Peroxide,
Ozone, Sulfur Dioxide, and Mineral Acids must be stored
in a freezer (less than 0°C; 32°F). However, because
temperatures in storage rooms can vary, Honeywell
Analytics recommends that all Chemcassettes
stored in a freezer, except Chlorine/Chlorine Dioxide,
Chlorine/Oxidizers, Phosgene and XP Phosgene must
be stored at 4°C to 25°C (40°F to 77°F).
Each Chemcassette
at time of manufacture. Under no circumstances should
a Chemcassette
be used beyond the expiration date.
®
Equally important, never remove a Chemcassette
its protective packaging until you are ready to install it
in your instrument. Exposure to light, ambient air, and
body oils may cause the Chemcassette
of its sensitivity. For more information on Chemcassette
care, refer to package instructions.
Note:
Chemcassettes
may be ordered on a 12-month blanket
®
purchase order; fresh Chemcassettes
on a scheduled basis. Consult Honeywell Analytics for
details.
3-3
®
. See
Section 4.9
®
detection systems will maintain
®
is stamped with an expiration date
®
to lose some
®
will be shipped
®
for
be
®
from
®
®

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents