Table 3-1. Network Device Requirements with Applicable Port Settings
Network Device
100 Mbps
• Switch
Auto-
Harmony Access Point
Negotiate
Controller
* The Switch port setting must be configured using the Config Tool. See Device
Configuration on page 4-7.
Note
All Switch to Switch connections need to operate at 100 Mbps FULL duplex. Acceptable port
settings are 100 Full to 100 Full or Auto-Negotiate to Auto-Negotiate.
Specific Network
Table 3-2 lists the current Clinical Network infrastructure devices and the speed and duplex
Device Settings
settings they support in the Information Center network.
Device
HP2524 Switch
Cisco 1900 Switch
8 Port Switch
J3300 10 Mbps Repeater
J3300 + J2606 10 Mbps
Translator Pair
100 Mbps Media Translator
(E-100BTX-FX-04)
100 Mbps Media Translator
(E-100BTX-FX-05)
* this device only supports 100Mbps operation
Drawing the
The final step is to draw the Patient Care Network using the devices, components, and
Design
cabling. Figure 1-6 illustrates a typical schematic design. Your final design, however, should
show all relevant information (Device Names, Locations, Cable Types and Lengths,
etc.).
Design
This section describes the guidelines to be followed when designing a Clinical Network
Guidelines
system. Two examples are given: one is a single switch network and the other is a multiple
switch network. These examples describe hardware design only.
The basic principles in designing the network design are:
Device NIC Setting
Auto-Negotiate or
100 Mbps Full Duplex
Auto-Negotiate
Table 3-2. Network Device Speed & Duplex Settings
Port(s)
Auto-Negotiate
1-24 (UTP)
yes
25-26 (fiber)
1-24 (UTP)
A (fiber)
B (UTP)
yes*
All
yes
yes*
yes*
Operational Speed
and Duplex Setting
Auto-Negotiate or
100 Mbps Full Duplex
100 Mbps Full Duplex
10Mbps Half
100 Mbps Half
yes
yes
yes
yes
yes
Network Design
Acceptable Switch
Port Setting*
Auto-Negotiate or
100 Mbps Full Duplex
Auto-Negotiate
100 Mbps Full
yes
yes
yes
yes
yes
yes