Network Design
Note
This manual does not describe the design and implementation of Clinical Network systems in
detail because each clinical environment requires careful analysis by an experienced network
designer. Philips Service Providers are specially trained to assist customers in reviewing their
clinical requirements and designing a Clinical Network system that meets those requirements
and will be supported by Philips Medical Systems.
The general concepts and capabilities of Network components and systems given here should
only be used as a reference for understanding the final design. Two examples are provided to
illustrate the design process.
Consult a Philips Service Provider for Clinical Network system design advice.
Connecting
Review the locations of the network devices and select the repeaters, media translators, cable
Devices
types and lengths required to interconnect them. (See Network Connections on page 4-45).
Table 3-1 summarizes the Clinical Network devices and the settings required for connection.
Start with device and switch selection, and follow across to determine what speed and duplex
settings are needed to make the connection. Then, determine which connection type is
appropriate. Table 3-2 shows the current hardware and the port settings they are capable of
supporting.
Note
The network connections for these devices are described in detail in Network
Connections on page 4-45.
Device Requirements with Applicable Port Settings
The following table represents the devices supported in the clinical network, the device
settings, operational speed and duplex requirements, and the acceptable switch port settings.
Table 3-1. Network Device Requirements with Applicable Port Settings
Network Device
10 Mbps
• Wired Patient Monitors
• Network Printers
• RangeLAN2 Access
Points
• Harmony Access Points &
Remote Power Supply
100 Mbps
• Information Centers
• Clients
100 Mbps
• Database Server
• Small Database Server
• Application Server
3-12
Site Planning
Device NIC Setting
10 Mbps Half Duplex
100 Mbps HALF Duplex
100 Mbps FULL Duplex
Operational Speed
and Duplex Setting
10 Mbps Half Duplex
100 Mbps HALF Duplex
100 Mbps FULL Duplex
Acceptable Switch
Port Setting*
Auto-Negotiate or
10 Mbps Half Duplex
100 Mbps HALF
Duplex
100 Mbps FULL
Duplex