Download  Print this page

Tests Recommended When Performing; Installation; Repair; Preventive Maintenance - Philips IntelliVue MP70 Service Manual

Patient monitor.
Hide thumbs

Advertisement

Tests Recommended When Performing...

Tests Recommended When Performing...

Installation

Service Event
(When performing...
Installation of monitor with no display connected
to the VGA output
Installation of monitor with a display connected
to the VGA output

Repair

Service Event
(When performing...
Repairs of M3015A
Repairs where the monitor has been damaged by
impact
Repairs where the power supply is replaced
All other IntelliVue Monitoring System repairs

Preventive Maintenance

Perform preventive maintenance tests:
• NBP calibration
• Sidestream CO
• Pump and scrubber replacement.

Performance Verifications

Perform all safety, accuracy and performance test procedures listed in the following sections. If a
particular measurement is in question, perform the measurement performance test only.

Upgrades

Service Event
(When performing...
Hardware and software upgrades
calibration
2
3 Testing and Maintenance
Test Blocks Required
...Complete these tests)
Perform Visual and Power On Test Blocks
Perform Visual, Power On and Safety (1) Test
Blocks
Test Blocks Required
...Complete these tests)
Perform Power On and M3015A tests
Perform Power On and Safety (2) and (3) Test
Blocks
Perform Safety (2) Test Block
Perform Power On Test Block
Test Blocks Required
...Complete these tests)
Perform Power On Test Block unless otherwise
specified in the Upgrade Installation Notes shipped
with the upgrade.
35