9 Anesthetic Gas Module
AWRR
ETCO
2
IMCO
2
inN
O
2
inO
2
et SEV
in SEV
et DES
in DES
Halogthane, Enflurane, Isoflurane
et
in

Alarm Delay:

10 seconds if no automatic zero calibration occurs within that time.

Apnea Alarm:

Delay Range:
Criterion
Alarm:

INOP Alarms

INOP alarms are triggered if:
• The Philips M1026A Anesthetic Gas Module is disconnected or switched off.
• The equipment malfunctions.
• The Agent-ID malfunctions.
• Zero calibration has failed.
• Zero calibration is in progress.
• The gas sample tube is occluded, or the water trap is full.
• The Philips M1026A Anesthetic Gas Module is unable to measure.
• Gas contaminant is detected.
• Agent mixture detected.
• Anesthetic agent detected but not selected.
• The module is in warm-up mode.
• No breath detected.
• The Anesthetic Gas Module is incompatible with the monitor
170
$JHQW
10 - 60 rpm
20 - 76 mmHg
2 - 20 mmHg
0 - 82 vol%
19-100 vol%
0.1 - 9.0 vol%
0.1 - 9.0 vol%
0.2 - 20.0 vol%
0.2 - 20.0 vol%
0.1 - 7.5 vol%
0.1 - 7.5 vol%
10 - 40 seconds
No detected breath within the adjusted delay time
Within 2 seconds after this criterion is met, if no automatic zero
occurs
+LJK 5DQJH
0 - 59 rpm
10 - 75 mmHg
none
none
18 - 99 vol%
0.0 - 8.9 vol%
0.0 - 8.9 vol%
0.0 - 19.8 vol%
0.0 - 19.8 vol%
0.0 - 7.4 vol%
0.0 - 7.4 vol%