Calibration and Electrical Safety Checks

Mechanical Inspection

Calibration and Electrical Safety Checks

Disposal

140
Carry out a mechanical inspection of the monitor.
Check transducers, cables, connectors and other accessories for cracks
and defects.
Carry out the Quick Test as described on page 151.
Carry out the Parameter Test as described on page 152.
To carry out a mechanical inspection of the monitor:
Make sure all exposed screws are tight.
Check the external cables and housings for splits, cracks or signs of
twisting.
Replace any cables that show serious damage.
You must calibrate the NIBP function of your monitor every 12 months.
Please refer to the Installation and Service Guide for your monitor for
details.
Perform electrical safety tests as described in the Installation and Service
Guide for you monitor (e.g., after repairs and upgrades).
Arrangements for the disposal of your monitor at the end of its working
life should conform to your country's laws regarding the disposition of
equipment containing electrical parts.