Replacing Scales; Replacing Used Oil/Fresh Oil/ Uv Dye Scale - Bosch ACS 751 Repair Instructions

A/c service unit
Hide thumbs Also See for ACS 751:
Table of Contents

Advertisement

en
18 | ACS 751/651 | Scales
Fig. 14:
Attaching the calibrating weight to the
refrigerant cylinder
14. Select "Place Standard Calibration Weight".
15. Enter the value of the calibrating weight.
i
The actual value is embossed on the calibrating
weight.
16. Select "Input Standard Weight".
17. Select "Calibrate". If you wish to abort the calibra-
tion, press c.
? The calibration process starts. Once the calibra-
tion is complete, the message
complete
is displayed.
"
You have successfully calibrated the scale for the
refrigerant tank.
|
1 689 975 223
2013-10-21
5.2
5.2.1
Fig. 15:
1. Switch off the ACS 751/651.
2. Disconnect the mains connection.
3. Open the front service hatch and disconnect the oil
and UV dye bottles.
4. Open the front panel and top panel (Chap. 6.1,6.2).
5. Disconnect the connections from the scale to the I/O
Calibration
board. .
6. Undo the screws on the top and bottom of the re-
taining plate for the scale.
7. Place the new scale into position.
8. Tighten the screws on the retaining plate.
9. Restore the I/O connections to the scale.
10. Close the front panel.
11. Connect the power cable to the mains supply.
12. Switch on the mains supply.
13. Turn on the power switch.
14. In the main menu, select "ACS Settings".
15. Perform calibration with the standard weight.

Replacing scales

Replacing used oil/fresh oil/ UV dye scale

Scales assembly
Robert Bosch GmbH

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Acs 651

Table of Contents