Please read this manual carefully before starting to use the instrument. address MIRIS AB Kungsgatan 115 Miris Ultrasonic Processor is used to prepare milk samples for 753 18 Uppsala Sweden macronutrient analysis using Miris HMA™ - Human Milk Analyzer. www.MirisSolutions.com...
TABLE OF CONTENTS PREFACE ............................3 TABLE OF CONTENT ........................4 IMPORTANT USER INFORMATION ....................5 INTENDED USE OF THE MIRIS ULTRASONIC PROCESSOR ........... 5 WARRANTIES AND DUTIES ..................... 5 SAFETY INFORMATION ......................6 MIRIS ULTRASONIC PROCESSOR USER GUIDE ................7 Chapter 1 INSTALLATION ......................
IMPORTANT USER INFORMATION INTENDED USE OF THE MIRIS ULTRASONIC PROCESSOR Miris Ultrasonic Processor reduces the particle size distribution in milk. This device is intended for use in laboratories by laboratory personnel. WARRANTIES AND DUTIES By operating this Miris Ultrasonic Processor, the USER and MIRIS agree to the following responsibilities, which constitute contractual warranties and conditions between MIRIS and USER for the maximum benefit and usefulness of the Miris Ultrasonic Processor.
SAFETY INFORMATION When mounting the probe, always clamp the converter housing. Never clamp the probe. Make sure the Ultrasonic Processor is properly grounded via a 3-prong outlet. High voltage is present in the power supply. Do not remove the cover. Service can only be performed by an authorized service technician.
MIRIS ULTRASONIC PROCESSOR USER GUIDE Chapter 1 INSTALLATION Figure 1. The Miris Ultrasonic Processor. 1. Sound reducing cabinet 2. Display and keypad Miris Ultrasonic Processor User Manual INSTALLATION | 7...
Page 8
16. Amplitude ELECTRICAL REQUIREMENTS The Miris Ultrasonic Processor requires a fused, single phase 3-terminal grounding type elec- trical outlet capable of supplying 50/60 Hz at 100 Volts, 115 Volts, 220 Volts, or 240 Volts, depending on the voltage option selected. For power requirements, check the label on the back of the unit.
MIRIS ULTRASONIC PROCESSOR USER GUIDE Chapter 2 SETTINGS TIME Processing time for human milk samples is 1.5 s/ml milk. Calculate sample processing time in seconds by multiplying sample volume in ml by 1.5 Set sample processing time by pressing [Timer]...
Set the processing time (1.5 s/ml) according to sample volume Make sure that the Miris Ultrasonic Processor probe is clean and that the tip is evenly polished (For instructions on human milk preparation and handling, see the Miris HMA™ User Manual)
Page 11
After each use of the Miris Ultrasonic Processor, wipe the probe with a tissue Cleaning At the end of the day, wipe the probe and all contaminated areas with a cloth dampened with Miris CLEANER™ Full instructions in Chapter 4...
Use solution within 3 months of preparation. Environment and health: Small spillages may be flushed away with plenty of water to a drain or sewer. Miris Cleaner™ is biodegradable. Ready biodegradability by OECD 301E (ISO method 7287 - 1986(E)). Non-hazardous. In case of contact with eyes, rinse immediately with plenty of clean, flowing water.
Page 13
PROBE CLEANING Wipe off the probe after every milk sample using a tissue. Clean the probe with a cloth dampened with Miris Cleaner™ after finishing for the day. Polish the tip (the flat end) of the probe with an emery cloth after each use.
Chapter 5 TROUBLESHOOTING AND PROBLEM SOLVING If any problem occur that you are unable to solve by referring to this manual, please contact your local distributor or Miris. When doing so, please include the serial number of your instrument. The serial number is printed on the label placed at the back of the Miris Ultrasonic Processor.
Time, particles [1]. The energy output from the Miris Ultrasonic Processor temperature and gravitation are some parameters that affect is approximately 20 J/s per ml of milk.