To install the HP E1433A

Caution
To protect circuits from static discharge, observe anti-static techniques
whenever handling the HP E1433A 8-Channel 196 kSa/s Digitizer plus DSP
Module.
1
Set up your VXI mainframe. See the installation guide for your mainframe.
2
Select a slot in the VXI mainframe for the HP E1433A module.
The HP E1433A module's local bus receives ECL-level data from the
module immediately to its left and outputs ECL-level data to the module
immediately to its right. Every module using the local bus is keyed to
prevent two modules from fitting next to each other unless they are
compatible. If you will be using the local bus, select adjacent slots
immediately to the left of the data-receiving module. The local bus can
support up to four HP E1433A modules at full span at real time data rates.
If the VXI Bus is used, maximum data rates will be reduced but the
module can be placed in any available slot.
3
Using a small screwdriver or similar tool, set the logical address configuration
switch on the HP E1433A.
(See the illustration on the next page.) Each module in the system must
have a unique logical address. The factory default setting is 0000 1000
(8). If an HP-IB command module will be controlling the HP E1433A
module, select an address that is a multiple of 8. If your VXI system
dynamically configures logical addresses, set the switch to 255.