GE TONOPORT V Service Manual page 34

Ambulatory blood pressure system
Hide thumbs

Advertisement

11-2
Checkout Procedure:
Voltages
Setpoint Value (V)
TP2
1.25 +/-0.1
TP3
1.25 +/-0.1
TP7
3.06 +/-0.1
TP8
3.30 +/-0.1
TP9
-3.25 +/-0.1
TP10
3.30 +/-0.1
Time for pumping up (to 170 mmHg with a container of 0.5 L)
Setpoint Value (sec)
t
<25
Pressure channel (tolerance
Displayed Pressure Value (mmHg)
50 +/-20 steps
50 mmHg
100 mmHg
150 mmHg
200 mmHg
250 mmHg
Leakage test
3 mmHg/minute?
Overpressure test passed?
Time limit test passed?
Visual inspection passed?
Checkout OK?
Date:..............................
TONOPORT V
2001589-094
Actual Value (V)
Actual Value (sec)
2 mmHg)
o Yes
o No
o Yes
o No
o Yes
o No
o Yes
o No
o Yes
o No
Name and signature of test person:
............................................................................
Remarks
Remarks
Remarks
Revision D