Sony DXC-390 Instructions For Use Manual page 3

3ccd color video camera
Hide thumbs Also See for DXC-390:
Table of Contents

Advertisement

Available languages

Available languages

Important safeguards/notices for use in the medical
environments
1. All the equipments connected to this unit shall be certified
according to Standard IEC60601-1, IEC60950, IEC60065
or other IEC/ISO Standards applicable to the equipments.
2. When this unit is used together with other equipment in
the patient area*, the equipment shall be either powered
by an isolation transformer or connected via an additional
protective earth terminal to system ground unless it is
certified according to Standard IEC60601-1.
* Patient Area
3. The leakage current could increase when connected to
other equipment.
4. This equipment generates, uses, and can radiate
frequency energy. If it is not installed and used in
accordance with the instruction manual, it may cause
interference to other equipment. If this unit causes
interference (which can be determined by unplugging the
power cord from the unit), try these measures: Relocate
the unit with respect to the susceptible equipment. Plug
this unit and the susceptible equipment into different
branch circuit. Consult your dealer.
(According to Standard EN60601-1-2 and CISPR11, Class
B, Group 1)
Caution
When you dispose of the unit or accessories, you must
obey the law in the relative area or country and the
regulation in the relative hospital.
GB
3
(GB)

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dxc-390p

Table of Contents