C.2.3 200-299 - Siemens SITRANS F MAG 8000 Standard Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS F MAG 8000 Standard:
Table of Contents

Advertisement

C.2.3
200-299
Param‐
Meter
Display
eter ID
version
view
200
All
O4
201
All
-
202
All
-
203
All
O4
204
All
-
205
All
-
206
All
-
SITRANS MAG 8000
Operating Instructions, 11/2018, A5E03071515-AE
Parameter/data type
Fault status
Alarm configuration list
Date of fault log reset
Non optimal measure time, in‐
cluding the time of device run‐
ning with insulation test fault,
coil current fault, amplifier
fault, database checksum
fault, flow overload fault, or
empty pipe fault.
Reset the fault log and faults
Call up acknowledge
Battery alarm level
Factory settings
Fixed parameter or meter data that are not changeable
No faults
254 = Alarm 2 to 8 enabled 0 to 8191, See 200
PS3 production date and
time
0
No
No
10%
Appendix
C.2 Parameter lists
Data range
0 to 8191, binary presented with
information 1 for bit 0
1: Insulation error
2: Coil current error
3: Amplifier overload
4: Database checksum error
5: Low power warning
6: Flow overflow warning
7: Pulse A overload warning
8: Pulse B overload warning
9: Consumption interval warning
10/L: Leakage warning
11/E: Empty pipe warning
12/C: Low impedance (high con‐
ductivity) warning
13/d: High flow limit warning
14/A: Reverse flow warning
If no error/warning is pending,
the message is "No faults"
If no alarm is enabled, the mes‐
sage is "None enabled"
year-month-day T hours:mi‐
nutes:seconds
Yes / No
Value is reset to "No" when up‐
loading device parameters to pc
after command execution
Yes / No
Value is reset to "No" when up‐
loading device parameters to pc
after command execution
0 to 100%
139