C.2 Parameter lists

Flow rate unit (parameter 9)
BBL42/h
BBL42/d
C.2

Parameter lists

Parameter lists
MAG 8000 is delivered with factory settings that are not stored as default values. Because
defaults values are not present in the meter, an automatic return to factory values is not
possible.
The default settings are available at www.siemens.com/flow (www.siemens.com/flow).
Navigate to Tools & Downloads under MAG 8000.
Visible display information is indicated in the table by menu and index number. Remember to
enable displayed menus in parameter 130.
The abbreviations used in the display menu table are: Operator menu = O, Meter menu = M,
Service menu = Se, Data Logger menu = L, Statistic menu = St, Revenue menu = R.
Factory settings marked in
Note
Change of units
If the unit and factor is changed, all unit-related values become invalid.
It is recommended to reset all unit-related values.
C.2.1
1-99
Param‐
Meter
Display
eter ID
version
view
1
All
M1
2
All
-
3
All
M3
4
All
M4
5
All
-
136
italic cannot be edited.
Parameter/data type
Application identifier
Application location
Module type
Software version
Sensor size
Correction factor parameter 301
22643.31867
543439.6482
Factory settings
Fixed parameter or meter data that are not changeable
Identity
Location
MLFB dependent
Sensor related
Operating Instructions, 11/2018, A5E03071515-AE
Data range
Max. 15 characters.
Only numbers are visible in the
display (letters are shown as "_")
Max. 16 characters
MAG 8000 Basic
MAG 8000 Advanced
MAG 8000 CT Basic
MAG 8000 CT Advanced
MAG 8000 Irrigation
x.xx
Max. 12 characters.
DN 25 to 1200 (1" to 48")
SITRANS MAG 8000