Page 4
Table of Contents Chapter 5 – Composer Software ....................35 Installation ______________________________________________________________________________ 35 Overview ................................35 Downloading the Software and System Requirements ..................35 Installing the Software ............................35 Network Configuration ............................35 Composer Welcome Screen Orientation ______________________________________________________ 36 Using Composer Software __________________________________________________________________ 37 System Overview Screen ___________________________________________________________________ 37 Dashboard ................................
Page 5
Table of Contents GPIB and Ethernet Modbus Interface Converters (optional) _____________________________________ 60 Serial Modbus-to-GPIB Converter (Options 1052 and 0003) ................60 Serial Modbus-to-Ethernet Converter (Option 1056) ..................60 Programming Syntax for the F4T Controller ..................... 61 Programming Syntax for the F4T Controller, Data Map 3 ................62 Chapter 7 –...
12" from the rear of the chamber. ! CAUTION: This chamber is designed for operation in a conditioned laboratory environment. Operation above 30°C (85°F) or below 16°C (60°F) ambient room temperature is NOT recommended. TestEquity 155 Temperature Chamber Page 7...
F4T Limit Alarm High Limit Set Point. ! CAUTION: The F4T Controller has been properly configured by TestEquity to match the chamber’s system requirements and to perform optimally over a wide range of operating conditions. Improper modifications to these setup values can result in erratic performance and unreliable operation.
Page 9
You are responsible for providing thermal protection devices to your test sample. ! CAUTION: To prevent damage to your test sample and the chamber’s compressor, do not exceed the live load rating of the chamber TestEquity 155 Temperature Chamber Page 9...
Chapter 2 – Installation Chapter 2 – Installation Unpacking Inspect the shipping container for any signs of visible damage. Notify the carrier and TestEquity immediately if there are signs of shipping damage. 1. Cut the bands that hold the packaging together.
After you reverse the door, see “How to inspect the door seal” in the Maintenance chapter of this manual to make sure the hinges and door latch are adjusted correctly. Page 12 TestEquity 155 Temperature Chamber...
Air Control key. The default condition is with the controller operating in Part Temperature Control mode (no green check mark in the Air Control box). See more information in the section entitled “Part Control vs. Air Temperature Control”. TestEquity 155 Temperature Chamber Page 13...
The port plug must be inserted with the gray silicone surfaces facing the inside and outside of the chamber. Port plugs should be considered expendable and be replaced when they no longer provide a good seal. Page 14 TestEquity 155 Temperature Chamber...
Make sure that all connectors, wiring, pc boards, and auxiliary components can withstand the temperature extremes that they will be subjected to. In some cases, these components may not be able to last after repeated tests and should be considered expendable. TestEquity 155 Temperature Chamber Page 15...
This chapter is designed to give you a better understanding of the structure and navigation of the F4T menus as viewed from the front panel. Understanding F4T Menus The graphic below illustrates at a high level the structure of the F4T menus. TestEquity 155 Temperature Chamber Page 17...
Next Page: Home screen has been setup to display multiple pages (loops). The left and right arrows on each side of the home screen provides navigation from one to the other (see “Personalizing the Home Screen” section). ⑤ Output Widget Bar: Function keys and their output status (ON/OFF). Page 18 TestEquity 155 Temperature Chamber...
• Personalize: The home screen layout is modified in this menu. This has been configured by TestEquity to only show if a user is logged in with a password. • Help: (not shown, appears when display is scrolled down in this pictured configuration) Displays information about the controller such as part number and software revision.
• Bksp (backspace) will clear the last digit in the Closed Loop Set Point box. • Enter will enter the values in the Closed-Loop Set Point box and return you to the Home Screen. • Cancel will return you to the Home Screen without changing the Set Point. Page 20 TestEquity 155 Temperature Chamber...
The buttons can be pressed to toggle the functions ON/OFF in Static Set Point (manual) mode or while a profile is running. Below is a summary of the button functions assigned by TestEquity. Power Button The Power button on the F4T Controller turns the chamber ON/OFF.
The screen which follows will provide access to the Low Limit Set Point and High Limit Set Point. Pressing the Low Limit or High Limit Set Point value will take you to a numeric key screen where you can enter new values. Page 22 TestEquity 155 Temperature Chamber...
Clear button is pressed. Pull the status bar down button to view the alarm message in the Error tab. Press Clear to clear the alarm after the chamber temperature is within safe limits. TestEquity 155 Temperature Chamber Page 23...
Part Temperature while minimizing lag time. The controller is configured by TestEquity to allow the Air Temperature to deviate up to +10°C or –10°C from the Part Temperature set point. This means if the set point is 85°C, the Air Temperature could go as high as 95°C if the Part Temperature lagged sufficiently during a heat...
When configured for Part Control (Air Control button has no green check mark), the F4T will attempt to control the temperature at the Part Sensor. In this mode, the Part Sensor must always be attached to your test sample. Failure to do this will result in erratic temperature control. TestEquity 155 Temperature Chamber Page 25...
The F4T is configured by TestEquity to allow the Air Temperature to deviate no more than +10°C or –10°C from the Part Temperature set point. This means if the set point is 85°C, the Air Temperature could go as high as 95°C if the Part Temperature lagged sufficiently during a heat...
This function is located in the Temperature Loop Operational Parameter screen. Pressing the vertical ellipsis or anywhere within the outlined box shown below will provide access to the Loop Operational Parameter screen. Simple Set Point Enable condition is persistent upon power recycle. TestEquity 155 Temperature Chamber Page 27...
This menu indicates which data points have been selected. Select Data Points You can select the parameters which get logged. TestEquity has pre-configured the controller to log the Set Point, Inner Process Value (air temperature sensor), and Outer Process Value (part temperature sensor).
2. Press the Data Logging button. 3. Press the Data Log File Transfer button. 4. Select Auto Transfer Type: TFTP, Samba, or USB. NOTE: If TFTP or Samba is selected above, the server must be configured. Page 30 TestEquity 155 Temperature Chamber...
To access trend charts, press the Menu button and then the Trending button. Press the Actions button for Trend Chart 1. This chart has already been set up by TestEquity to display the Air Sensor, Part Sensor, and Set Point.
6. A progress bar will appear and then indicate 100% when done. Importing a configuration file is accomplished similar to the profile import except you select Configuration and then selecting the desired configuration file. Page 32 TestEquity 155 Temperature Chamber...
• From any screen, press the Menu button. • Press Personalize. • Press Basic Personalization. • Select the desired Home Layout by pressing Main Top, Main Center, or Main Bottom. • Press Page 1. • Select “2 Content Blocks”. TestEquity 155 Temperature Chamber Page 33...
PID tuning parameters. ! CAUTION: The F4T PID tuning values have been properly configured by TestEquity to match the chamber’s system requirements and to perform optimally over a wide range of operating conditions. Improper modifications to these values can result in erratic performance and unreliable operation.
Overview Watlow Composer Software can be used to create, edit, and archive profiles. It can also be used to configure the controller. Since the controller is already configured by TestEquity for use in your chamber, configuration instructions are not provided.
⑦ Help: Allows you to change automatic software update settings, as well as initiate an immediate check for updates (internet connection is required). Clicking on About will display technical support contact information and the current firmware versions of the installed software and installed modules. Page 36 TestEquity 155 Temperature Chamber...
• User with Password - password required with limited access based on settings. • Maintenance User Password - password required with limited access based on settings. • Admin Password - password required with full unlimited access. TestEquity 155 Temperature Chamber Page 37...
Page 38
• Setup, controls access to the pluggable module configuration and the function block diagram. The image below shows the security levels as configured by TestEquity. You must contact TestEquity to get the passwords for your chamber. After making all the desired security settings, ensure that the security enabled radio button (top left in the graphic above) is checked.
2. Use the save as dialog to select the destination folder for the image. 3. Enter the desired filename. 4. Click Save. Note: The system image filename will always have the extension wsi for Watlow System Image and cannot be changed. TestEquity 155 Temperature Chamber Page 39...
3. Double-click the desired filename or single-click the filename and then click the Open button. 4. After a system image is imported, we suggest you cycle power to the F4T Controller and restart Composer software. Page 40 TestEquity 155 Temperature Chamber...
The Pluggable Module Screen shows what modules are installed. TestEquity has already installed and configured the correct modules for your chamber. Since this is not intended to be an ordinary user function, no further instructions will be covered in this manual.
Data Logging See Data Logging in Chapter 2 for an explanation of these settings. TestEquity suggest you select the minimum number of relevant parameters to log. This will result in a smaller file size. In the example below for a temperature chamber, only Temperature Set Point and Temperature Process Value (the actual chamber temperature) were selected.
Password Protect a Profile: Avoid unwanted and inadvertent changes to a profile with password protection. Changing or Removing a Password from a Profile: When password protection is no longer needed it can be easily removed. TestEquity 155 Temperature Chamber Page 43...
③ Step Detail • Shows all available selections for the selected step while also allowing each to be modified. • You have the ability to give the profile a name as well as apply password protection. Page 44 TestEquity 155 Temperature Chamber...
When multiple conditions are specified, the profile will not proceed until all the conditions are satisfied at the same time. • END: Sets the temperature set points, and event output conditions when the profile ends. See page 16 for more information. TestEquity 155 Temperature Chamber Page 45...
Page 46
• On: the step sets the event output on or true. • Off: the step sets the event output off or false. • Unchanged: the step does not set the event output; it remains in whatever state was previously set. Page 46 TestEquity 155 Temperature Chamber...
To load a profile previously saved on the computer in to controller memory: 1. Click the Load button. 2. Use the open dialog to locate and select the desired profile file. 3. Click Open. TestEquity 155 Temperature Chamber Page 47...
To insert a new step ahead of an existing step: • Right-click the step and choose Insert Step Before in the pop-up menu. • Click the step and then click the Insert Before button at the bottom of the step list. Page 48 TestEquity 155 Temperature Chamber...
6. Click OK. To remove a profile’s password: 1. Select the profile by clicking it in the profile list. 2. Enter the password and click OK. 3. Click the Password button. 4. Click the Remove button. TestEquity 155 Temperature Chamber Page 49...
Data Map, the user can switch Modbus registers from the comprehensive listing of F4T registers. Map 1 = F4T registers (default) Map 2 = Not applicable to this Model 155 or controllers with Cascade Control (Part Temperature Control) Map 3 = For controllers with Cascade Control including this Model 155. Limited set of F4 compatible registers (for compatibility with software written for the F4).
• Click the radio button identified as Use the following IP address. • Change the IP address to 192.168.0.2, Subnet mask to 255.255.255.0 (this assumes the F4T is configured for 192.168.0.222 which is how they ship from TestEquity, you can choose a different IP address).
• If there is no Router between the Computer and the F4T, then when Composer Software allows you to “Connect to a System to Configure it” the F4T Name/IP Address should automatically be displayed. Select this F4T Name/IP Address to use Composer with this F4T (this chamber that is). TestEquity 155 Temperature Chamber Page 53...
Data Map, the user can switch Modbus registers from the comprehensive listing of F4T registers. Map 1 = F4T registers (default) Map 2 = Not applicable to this Model 155 or controllers with Cascade Control (Part Temperature Control) Map 3 = For controllers with Cascade Control including this Model 155. Limited set of F4 compatible registers (for compatibility with software written for the F4).
3. Select the Ethernet (for LAN) or Modbus (for RS-232) communications channel 4. Scroll the screen down to find Data Map and select 1 (for F4T) or 3 (for legacy F4 Compatibility Mode). 5. Recycle the power to put the new settings into effect. TestEquity 155 Temperature Chamber Page 55...
Manual Ramp Parameters (Ramp to Set Point without a profile) Ramp Action (Temperature) 2794 Off = 62 Startup = 88 Set Point = 85 Both = 13 Ramp Scale (Temperature) 2796 °/Minutes = 57 °/Hours = 39 Ramp Rate (Temperature) 2798 Page 56 TestEquity 155 Temperature Chamber...
Page 57
0 to 40 Current Step 16590 (R) 0 to 50 Step Type 16592 (R) Soak = 87 Wait For = 1542 Instant Change = 1927 Ramp Time = 1928 Ramp Rate = 81 End = 27 TestEquity 155 Temperature Chamber Page 57...
Serial Modbus-to-GPIB Converter (Options 1052 and 0003) Your chamber may have been ordered with a GPIB Interface option. GPIB communications is achieved through an ICS Electronics 4899A (external, TestEquity Option 1052) or 4809A (internal, TestEquity Option 0003) GPIB-to-Modbus Interface Converter. The ICS 4899A converts GPIB commands to serial Modbus commands that are transmitted over RS-232.
Read Closed Loop Set Point RF? 2810 (instantaneous temperature set point during a ramp or profile) Turn (Event 1) ON W 16594, 63 Event 1 is “Power” Turn (Event 1) OFF W 16594, 62 TestEquity 155 Temperature Chamber Page 61...
Data = ASCII data written as 16-bit decimal value Examples: Write Temperature Set Point W 300, value Read Chamber Temperature R? 100, 1 Event 1 ON W, 2000, 1 Event 1 OFF W, 2000, 0 Page 62 TestEquity 155 Temperature Chamber...
How can I modify the chamber to cool faster or colder? Unfortunately, there is nothing you can do to improve upon the designed-in performance. TestEquity does NOT recommend using CO or LN in this chamber to achieve colder or faster cooling due to reliability and safety considerations, so it is NOT an available option.
Page 64
This will ensure that nothing has been damaged and that a leak has not developed. Can/Should I put a filter in front of the condenser air inlet? No, TestEquity does not recommend this. Just follow the maintenance procedures and clean the condenser fins periodically.
Operation at higher ambient temperatures will result in decreased cooling performance. Low end limit derates to -38°C when operating above 27°C (80°F) ambient. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended. TestEquity 155 Temperature Chamber Page 65...
• Listen for abnormal noise or vibration. Every 6 Months • Inspect the door seal. • Clean the condenser. • Inspect the electrical/refrigeration compartment. • Verify the chamber performance. Every 12 Months • Verify the calibration. TestEquity 155 Temperature Chamber Page 67...
NOTE: You may need to clean the condenser more frequently if the chamber is in a dusty environment. You may be able to clean the condenser less frequently if the chamber is in a very clean environment. Page 68 TestEquity 155 Temperature Chamber...
4. Inspect for oil around refrigeration valves, fittings and joints. This may be a sign of leaks. 5. Inspect for loose hardware and tighten as required. 6. Check for loose wires and burned insulation near terminals. 7. Inspect for signs of insect or rodent infestation. Yes, it does happen! TestEquity 155 Temperature Chamber Page 69...
Temperature Controller’s setup and tuning values have not been changed from the values as shipped from TestEquity. Also, the Limit Controller high limit must be set to over +85°C (+100°C would be fine), and the low limit set to –27°C. If the chamber fails any of these tests, it should be removed from service to prevent further damage until the cause of the problem is determined and resolved.
TestEquity recommends verifying the calibration before attempting to actually perform a calibration. The microprocessor-based instrumentation used in TestEquity chambers seldom goes out of calibration. If you try to calibrate the instrumentation before determining that calibration is necessary, you may make it worse if done incorrectly.
• To enter the Calibration menu, press the Menu button, then the Service button, then the Calibration button, then the Part Sensor Action button. Next, press Perform Field Calibration and follow the on-screen prompts. Page 72 TestEquity 155 Temperature Chamber...
Page 73
• To enter the Calibration menu, press the Menu button, then the Service button, then the Calibration button, then the Alarm Sensor Action button. Next, press Perform Field Calibration and follow the on-screen prompts. TestEquity 155 Temperature Chamber Page 73...
% cool throttle, providing substantial savings in energy consumption (patent pending control algorithm). At temperatures below –16°C, the compressor speed will remain at 100% while the % cool throttle continues to cycle SV1. Page 74 TestEquity 155 Temperature Chamber...
Page 75
The suction line cooling thermostatic expansion valve (TEV1) senses the suction line temperature and injects liquid refrigerant to cool the hot gas within safe limits. This above described refrigeration system is patented by TestEquity under U.S. Patent Number 10,126,032. TestEquity 155 Temperature Chamber...
If tuning PID control parameters in a) Control parameters may need to be temperature controller are as shipped from changed for your unique test conditions. TestEquity and only occurs in cool mode. b) Solenoid valve SV1 may be defective. Compressor If controller Compressor Output parameters a) Re-enter values as shipped from were changed.
Do NOT substitute any component. Do NOT substitute refrigerants. Improper repairs will void the warranty. These instructions are intended as guidelines for repairing TestEquity chambers. Details such as how to attach a gauge manifold are not covered. These are NOT do-it-yourself instructions!
Chapter 9 – Maintenance Replacement Parts Replacement parts are available from TestEquity. Parts are usually in-stock and ready for immediate shipment. Next-day delivery is always available. Major Electrical Parts Description Mfr Part No. Ref # Part # Qty UOM Heater, Air...
If the repair or replacement is performed in the FIRST YEAR from the date of invoice, TestEquity will also pay for the labor associated with the repair at the Purchaser's site, subject to TestEquity’s prior approval. During the SECOND and THIRD YEAR of the warranty period, Purchaser will be responsible for the installation and cost of installation of replacement or repaired parts.
Need help?
Do you have a question about the 155 and is the answer not in the manual?
Questions and answers