TANCO Steam Sterilizer
Operation Manual
Operation Documentation
Copyright © 2011 By Woson Medical Instrument Co., Ltd.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TANCO and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Woson TANCO

  • Page 1 WOSON Medical System Technical Publications ASS0003 REV- B TANCO Steam Sterilizer Operation Manual Operation Documentation Copyright © 2011 By Woson Medical Instrument Co., Ltd.
  • Page 2: Regulatory Requirement

    Steam Sterilizer Operation Manual Regulatory Requirement This product complies with regulatory requirements of the following European Directive 93/42/EEC concerning medical devices. Ningbo Jiangbei Woson Medical Instrument Co., Ltd No.25, Lane 300, Jinshan Road, Jiangbei District, Ningbo 315032, China Tel: 86-574-83022668 Fax:86-574-87639376 Website: www.woson.com.cn...
  • Page 3: Revision History

    Please verify that you are using the latest revision of this document. Information pertaining to this document is maintained on manufactory. If you need to know the latest revision, please contact your distributor, sales representative, or our service dept. TANCO Model REV-B Page 2 of 44...
  • Page 4: Regulatory Requirements

    The CE label affixed to the product testifies compliance to the Directive. The location of the CE marking is shown in this manual. In this manual there are present the CE certification and the Conformity. Check appendixes. TANCO Model REV-B Page 3 of 44...
  • Page 5: Table Of Contents

    Chapter 5 Panel and Functions ...................... 19 5.1 Controlling Panel ......................19 5.2 Menu ..........................20 5.3 Window of Sterilization Process ..................22 Chapter 6 Operation Process ......................24 6.1 Power On........................... 24 6.2 Distilled Water Adding ...................... 24 TANCO Model REV-B Page 4 of 44...
  • Page 6 Appendix 1 Instrument Which Need To Sterilize ................38 Appendix 2 Error Code List ......................39 Appendix 3 Electric And Hydraulic Drawings ................40 Hydraulic Drawing ........................40 Electric Drawing ........................41 Appendix 4 The Standards Of Testing ..................... 42 TANCO Model REV-B Page 5 of 44...
  • Page 7: Chapter 1 Introduction

    This sterilizer can be used for dental clinic, laboratory, surgical room, emergency room, ophthalmology, gynecology and steam, cosmetic hospital and so on, by doctors and professionals. 1.3 Contraindication There is no any contraindication of this equipment. TANCO Model REV-B Page 6 of 44...
  • Page 8: Chapter 2 Safety

    Be aware of electric shock hazards. Label Description Symbol for "SERIAL Symbol for NUMBER" "MANUFACTURER" Symbol for "CATALOGUE Symbol for "AUTHORISED NUMBER" REPRESENTATIVE IN THE EUROPEAN COMMUNITY Symbol for "DATE OF Symbol for "CAUTION" MANUFACTURE" TANCO Model REV-B Page 7 of 44...
  • Page 9: General Safety Recommendations

    Do not fill the caustic matter in the chamber. ❖ Use only high quality distilled water. ❖ Unplug the mains lead before inspecting or servicing the machine. ❖ Only an approved technician using original spare parts can carry out repairs and TANCO Model REV-B Page 8 of 44...
  • Page 10: Safty Parts

    Tray tong when removing articles from the chamber Control Part Part Name Function Temperature Sensor To measure temperature inside the chamber (Inside Chamber) Temperature Sensor To measure temperature of the heating ring TANCO Model REV-B Page 9 of 44...
  • Page 11: Operation Risk

    Please clean the chamber after every time use to avoid rudimental contamination left inside chamber. 2.5 Protection Device Device Name Function Plastic or fabric glove Useful during load and remove articles, avoid scald. TANCO Model REV-B Page 10 of 44...
  • Page 12: Chapter 3 Receiving And Installation

    Part Name Tray frame 1 piece Trays 3 sets Tray removal tong 1 piece Draining tube 1 piece Power supply cable 1 piece Door gasket 1 piece Operation manual 1 piece TANCO Model REV-B Page 11 of 44...
  • Page 13: Optional Accessories

    ❖ Do not put any stuff in front of the door, to avoid accident when open the door. ❖ Do not put any corrosive stuff near the sterilizer to avoid accident or risk. TANCO Model REV-B Page 12 of 44...
  • Page 14: Power Connection

    DO NOT PUT ANY HEAVY STUFF ON THE POWER WIRE TO AVOID DAMGE OF POWER WIRE DO NOT USE OTHER POWER WIRE TO AVOID DAMAGE OF STERILIZER DO NOT TRY TO ADD THE POWER WIRE TO AVOID ACCIDENT AND RISK TANCO Model REV-B Page 13 of 44...
  • Page 15: Chapter 4 Description And Specification

    2.Control Panel Operation button to control the process of unit 3.Filling Water Port Manually 4.Door Handle Open and close the door of sterilizer 4.2 Rear View of Sterilizer TANCO Model REV-B Page 14 of 44...
  • Page 16: Opened Vew Of Sterilizer

    To load the sterilized instrument 14. Pressure protector It ensures the door cannot be opened when chamber with pressure 15. Drainage Port For distilled water out 16. Drainage Port For used water out TANCO Model REV-B Page 15 of 44...
  • Page 17: Outside Size

    The board affording weight: 4000 N/m Noise :< 50db The maximum capacity of a plate:1000g The maximum duration of using loading test: 90mins. The maximum thermal radiation energy under the condition of 20℃~26℃:<2000J. TANCO Model REV-B Page 16 of 44...
  • Page 18 CAUTION WATER ADDED INTO MAIN WATER TANK MUST USE DISTILLED WATER! AND WATER TEMPERATURE MUST UNDER 40℃。 Used water tank volume: 8L (4.5L) 12L (4.5L) Test Method Vacuum Test B&D Test Helix Test TANCO Model REV-B Page 17 of 44...
  • Page 19: Sterilization Cycle

    The description identifies ranges of products and test loads. Non wrapped sterilized instruments are intended either for immediate use or for non NOTE 2 sterile storage, transport and application (e.g. to prevent cross infection). TANCO Model REV-B Page 18 of 44...
  • Page 20: Chapter 5 Panel And Functions

    CAUTION PLEASE CONTACT YOUR DISTRIBUTOR OR AUTHORIZED PERSON WHEN ERROR CODE DISPLAYED. TANCO Model REV-B Page 19 of 44...
  • Page 21: Menu

    One time vacuum, working temperature 121℃; 121℃ 1.1 bar 15min working pressure 2.1bar; sterilization cycle time 15 minutes; and drying time 17 minutes. Press Press button to confirm. button to choose the program, TANCO Model REV-B Page 20 of 44...
  • Page 22 = 09 correct value, and press to go next minute setting window. TIME SET Press button to choose the Minute = 29 correct value, and press to save and quit TANCO Model REV-B Page 21 of 44...
  • Page 23: Window Of Sterilization Process

    03:00 Second exhausting Third vacuum -0.32bar 86℃ 2.1bar 134℃ prssurize --:-- Sterilization 05:00 Third pressurizing Sterilizing 1.56bar 133℃ -0.23bar 97℃ Exhausting 05:00 Drying 3:45 Exhausting Drying -0.01bar 80℃ 0.00bar 85℃ Waiting --:-- Waiting TANCO Model REV-B Page 22 of 44...
  • Page 24 First pressurizing First exhausting 0.01bar 97℃ 2.1bar 134℃ Pressurize --:-- Sterilization 01:31 Third pressurizing sterilizing 0.56bar 105℃ -0.23bar 97℃ Exhausting 02:04 Drying 3:45 Exhausting Drying -0.01bar 80℃ 0.00bar 85℃ Waiting --:-- Waiting Finished TANCO Model REV-B Page 23 of 44...
  • Page 25: Chapter 6 Operation Process

    CAUTION USE DISTILLED WATER ONLY TO EXTEND THE LIFE TIME OF THE MACHINE. DO NOT TILT THE STERILIZER WHEN THE RESERVOIR IS FULL OF WATER. TANCO Model REV-B Page 24 of 44...
  • Page 26: Alarming If Used Water Reservoir Is Full

    Articles should be put on the trays with some gap between each articles so that the steam can be ventilated freely. Please use the attached handle tool to load instrument trays into chamber to avoid scalding .(pic 6-3) (pic 6-3) TANCO Model REV-B Page 25 of 44...
  • Page 27: Closing The Door

    Or you can push the door hardly in while you turn the door handle. If you still unsure the door has been closed properly, you can also adjust the door (refer to adjusting the sterilizer door) TANCO Model REV-B Page 26 of 44...
  • Page 28 COMPLETELY. IF THE DOOR HAS BEEN OPENED DURING THE CYCLE, THE STERILIZER WILL DISPLAY: DOOR IS OPEN DURING THE CYCLE! IF SO, PLEASE PRESS TO CANCEL THE ALARM, THEN CLOSE THE DOOR COMPLETELY AND RESTART. TANCO Model REV-B Page 27 of 44...
  • Page 29: Starting A Sterilization Program

    The buzzer make a sound means the total sterilization processes have been finished, then wait for the pressure down to “0” bar at the steam manometer on command front panel. CAUTION DO NOT PUT OR COVER ANY STUFF ON THE MACHINE TO KEEP HEAT VENTING WELL. TANCO Model REV-B Page 28 of 44...
  • Page 30: End Of The Sterilization Working Cycle

    If you will stop use it for a long time or for storage, please unplug the power cord. ATTENTION DURING THE STERILIZING, WE SUGGEST THAT YOU USE THE INDICATOR TAPE. PUT THEM IN THE CHAMBER IN ORDER TO ENSURE RELIABILITY OF STERILIZATION. TANCO Model REV-B Page 29 of 44...
  • Page 31: Abnormal Exiting

    Then skip to original state, remind you to open the door, open and close the door then to go to the next sterilization cycle. WARNING DO NOT TRY TO OPEN THE DOOR IF THE PRESSURE INSIDE CHAMBER DOES NOT SHOW "0 BAR". TANCO Model REV-B Page 30 of 44...
  • Page 32: Chapter 7 Essential Information

    …place the sterilizer in direct sunlight. ❖ …place the sterilizer on heat sensitive surfaces. ❖ …use inappropriate cleaning materials. ❖ …drop or abuse the sterilizer. ❖ …use in areas of risk associated with flammable materials or gases. TANCO Model REV-B Page 31 of 44...
  • Page 33: Chapter 8 Maintenance

    Refer to qualified personnel for servicing: Never use a wire brush, steel wool, abrasive material, or chloride-containing products to clean door and chamber assembly. “Caution hot surface. Avoid contact." ensure that the sterilizer is cooled down fully before cleaning to avoid burns. TANCO Model REV-B Page 32 of 44...
  • Page 34: Weekly Maintenance

    The kinds of impurity come from smeary dust on the instruments being sterilized or some calcification in the water. Keep cleaning of the inside chamber in order to make life-time of filter much longer; please TANCO Model REV-B Page 33 of 44...
  • Page 35: Monthly Maintenance

    BE SURE THAT USE DISTILLED WATER PROPERLY IN ORDER TO EXTEND THE STERILIZER LIFE-TIME. DO NOT RAVE ABOUT THE STERILIZER WHEN THE TANK HAS BEEN FILLED. DO NOT ADD MORE THAN 7 LITERS OF WATER TO THE RESERVOIR AFTER CLEANING. TANCO Model REV-B Page 34 of 44...
  • Page 36: Other Maintenance

    Attention: the inner edge of gasket may be ectropium during embedding it in the door groove, at this time; you’d better to tight it back to the groove by using the screwdriver carefully. TANCO Model REV-B Page 35 of 44...
  • Page 37: Chapter 9 Servicing By The Approved Technician

    Checking the hydraulic connections. Checking the safety thermostat. Cleaning the sterilization chamber. Cleaning the trays and the tray holder. Cleaning the reservoirs. Replacing the water filter. Replacing the air filter. Replacing the door gasket. TANCO Model REV-B Page 36 of 44...
  • Page 38: Chapter 10 Transportation And Storage

    The sterilizer package requirement should as followed: 1) Product can not over 3/4 volume of package 2) Product should be fix inside the package 3) Package bag should be higher than product 6mm TANCO Model REV-B Page 37 of 44...
  • Page 39: Appendix 1 Instrument Which Need To Sterilize

    4. Run appropriate sterilization program 5. Take out and store CAUTION STERILIZED INSTRUMENT SHOULD CHECK INSTRUMENT WITHOUT DAMAGE OR AVULSION. THE STERILIZED INSTRUMENT STILL WITH HIGH TEMPERATURE SHOULD NOT STACK, TO AVOID BAD HEAT RADIATION. TANCO Model REV-B Page 38 of 44...
  • Page 40: Appendix 2 Error Code List

    Pressure can not be exhausted Door opened during cycle Working over time Abnormal pressure Abnormal temperature Pressure and temperature can not match Reserved Vacuum failure or tube leaking Reserved Communication failure: Power interrupted during cycle Abnormally exit TANCO Model REV-B Page 39 of 44...
  • Page 41: Appendix 3 Electric And Hydraulic Drawings

    Air Outlet Valve Vacuum Pump Chamber Water Inlet Relief Valve Valve Vacuum Valve Distilled Water Drain Port Drying Air Inlet Valve Used Water Drain Port Pressure sensor Water Adding Port Door security System Temperature Sensor TANCO Model REV-B Page 40 of 44...
  • Page 42: Electric Drawing

    Inside Temperature Line No.4 (Red) Sensor 2 N (Black) Outside Temperature L (Red) Sensor 3 Line No.1 (Red) Output to Pressure To DB9 Port 220V Printer Sensor Electronic Filter Power Switch 12A Fuse Ground TANCO Model REV-B Page 41 of 44...
  • Page 43: Appendix 4 The Standards Of Testing

    Under the normal situation, sterilizer should indicate: a) Chamber temperature Indicator and display b) Chamber pressure c) Sterilizing work state d) Water level state e) Station of the opening or closing door TANCO Model REV-B Page 42 of 44...
  • Page 44 Any remaining water Dryness, solid load, double wrapped droplets on the inner side of the film of laminate pouch shall evaporate within 5 min. For load the moisture content shall not exceed 0,2 % TANCO Model REV-B Page 43 of 44...