Caution
Do not use delivery set or components that
have been used for prior testing for patient
infusion.
Before you use the delivery set (tubing, syringe,
etc.), make sure it is within the specified use
period of the manufacturer. Many sets are made
to be used only once.
Connect Drains to the Product
Figure 4 shows tubing connected to the rear-panel outlets
of the Product.
Figure 4. Drain Connections to the Product
Note
gir04.eps
When you connect drain tubing to the Product outlets:
•
Connect different drain tubes to each channel.
•
Do not connect the drain tubes together.
•
The drain tubes should not be allowed to rise more
than 10 cm (4 in) at any point above the height of the
inlet ports of the Product.
•
The discharge end of the drain tubes must not be
more than 10 cm (4 in) below the bottom of the
Product.
Connect Accessories
Accessories connect to any of the four USB "A" ports on
the rear-panel of the Product. Use a USB cable that is
less than 3 meters long.
When an accessory is connected to a Product
that is ON, some seconds are necessary before
the accessory is recognized.
Infusion Device Analyzer
Connect Accessories
Note
9
Need help?
Do you have a question about the IDA-5 and is the answer not in the manual?
Questions and answers