5. Adjust Knob
Used to increase (clockwise) or decrease (anticlockwise) the current
numerical value when setting parameters.
Also used to change characters when inputting a file name.
When saving or reading files, rotate the knob to choose a position to save a
file or choose a file to be read; press the knob to open a selected folder or
file.
6. Arrow Keys
Used to adjust digits in numeric values when using the knob to set
parameters.
Also used to position the cursor when inputting a file name.
When saving or reading files, the arrow keys are used to choose a position
to save a file or choose a file to be read.
7. CH1 Control Port/Output Button
The Output button on the left is used to turn on/off CH1 output.
The nominal output impedance of the BNC connector is 50 Ω.
When pressing Output the key backlight turns on, and the connector
outputs waveforms according to the current configuration of CH1.
CH2 Control Port/Output Button
The Output button on the right is used to turn on/off CH2 output.
The nominal output impedance of the BNC connector is 50 Ω.
When pressing Output, the backlight turns on, and the connector outputs
waveforms according to the current configuration of CH2.
!
CAUTION:
Over voltage protection of CH1 and CH2 will take effect once any of the following
conditions is met. When over voltage protection occurs, a message will be displayed
and the output is disabled.
•The absolute value of input voltage is higher than 11 V ± 0.5 V when the amplitude of
the generator is higher than or equal to 3.2Vpp or the DC offset is higher than or equal
to |2VDC|.
•The absolute value of input voltage is higher than 4 V ± 0.5 V when the amplitude of
the generator is lower than 3.2 Vpp or the DC offset is lower than |2 VDC|.
Choose Utility →
to turn on/off the function.
Quick Start 5