Influence Of Fluid Pressure; Design Fundamentals - Endress+Hauser Proline Promass 40E Technical Information

Coriolis flowmeter
Hide thumbs Also See for Proline Promass 40E:
Table of Contents

Advertisement

Proline Promass 40E

Influence of fluid pressure

Design fundamentals

Endress+Hauser
The table below shows the effect on accuracy of mass flow due to a difference between calibration
pressure and process pressure.
DN
[mm]
[in]
8
³⁄₈
15
½
25
1
40
50
2
80
3
o.r. = of reading
BaseAccu = base accuracy in % o.r.
BaseRepeat = base repeatability in % o.r.
MeasValue = measured value (in flow units consistent with the zero point stability value →  7)
ZeroPoint = zero point stability
Calculation of the maximum measured error depending on flowrate
Flowrate (in flow units consistent with the zero
point stability value →  7)
ZeroPoint
³
100
BaseAccu
ZeroPoint
<
100
BaseAccu
Calculation of the repeatability depending on flowrate
Flowrate (in flow units consistent with the zero
point stability value →  7)
½
ZeroPoint
³
100
BaseRepeat
½
ZeroPoint
<
100
BaseRepeat
Example for maximum measured error
E [%]
2.5
2.0
1.5
1.0
0.5
0
0
10
20
E = Error: Maximum measured error as % o.r.
Q = Flow rate as %
Promass E
[% o.r./bar]
no influence
no influence
no influence
no influence
–0.009
–0.020
Maximum measured error in % o.r.
± BaseAccu
A0021332
ZeroPoint
±
MeasValue
A002133
Repeatability in % o.r.
A0021335
ZeroPoint
±
½
MeasValue
A0021336
30
40
50
60
100
100
100 Q [%]
70
80
90
A0021339
A0021334
A0021340
A0021337
A0021279
9

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline Promass 40E and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents