Download  Print this page

Advertisement

P O W E R
T O
S A V E
A
L I F E
H E A R T S T A R T T R A I N E R
I N S T R U C T I O N S F O R U S E
M5085A
Edition 2

Advertisement

Table of Contents

   Also See for Philips HeartStart Trainer M5085A

   Summary of Contents for Philips HeartStart Trainer M5085A

  • Page 1 P O W E R S A V E L I F E H E A R T S T A R T T R A I N E R I N S T R U C T I O N S F O R U S E M5085A Edition 2...
  • Page 2 Intentionally blank.
  • Page 3 H E A R T S T A R T T R A I N E R M5085A I N S T R U C T I O N S F O R U S E Edition 2...
  • Page 4 Philips Medical Systems at (+49) 7031 463-1552 www.medical.philips.com/cms or your local distributor for information on Device Manufacturer revisions. The HeartStart Trainer is manufactured by Philips Medical Systems, Seattle, Washington, USA. Edition History Edition 2 Trademarks Publication date: June 2003...
  • Page 5: Table Of Contents

    Contents Intended Use ..................... Features ...................... Overview ....................HeartStart Trainer Kit Contents ............Optional Accessories ................Installing or Replacing the Batteries ............ Using the Manikin Adapters ..............Installing the External Manikin Adapter ........Installing the Internal Manikin Adapter ......... Installing or Replacing the Training Pads Cartridge ....Selecting a Training Scenario ...............
  • Page 6 Intentionally blank.
  • Page 7: Intended Use

    HeartStart Trainer M5085A Intended Use The HeartStart Trainer M5085A is designed to prepare emergency responders to use the HeartStart HS1 Defibrillators. The HS1 Defibrillators include the HeartStart OnSite Defibrillator M5066A, the HeartStart Defibrillator M5067A, and HeartStart Home Defibrillator M5068A. The HeartStart Trainer cannot be used to deliver defibrillation treatment.
  • Page 8: Overview

    Overview front view back view The HeartStart Trainer M5085A is designed to look like the HeartStart HS1 Defibrillator. However, some of the defibrillator’s physical features are either inactive in the Trainer or behave differently from their counterparts in the defibrillator, as noted below.
  • Page 9 blinks when the defibrillator is in standby mode, is on solid when the defibrillator is in use, and is off when the defibrillator needs attention.) ON/OFF BUTTON. You can also turn on the Trainer by pressing this green button. To turn off the Trainer, press the green button again and hold it down for one (1) second.
  • Page 10: Heartstart Trainer Kit Contents

    HeartStart Trainer Kit Contents The HeartStart Trainer M5085A comes with the following: • Trainer • Adult training pads cartridge • Carry case (does not include scissors) • Instructions for Use • External Manikin Adapter strip (1) • Quick Reference Optional Accessories •...
  • Page 11: Using The Manikin Adapters

    NOTE: Do not mix battery brands. Do not mix old and new batteries. Replace all batteries at the same time. If the Trainer is not to be used for some time, remove batteries to prevent possible leakage and corrosion. Using the Manikin Adapters The training pads should be used on a CPR training manikin equipped with a disposable External Manikin Adapter M5089A (A) or an Internal Manikin Adapter M5088A (B), to simulate the patient.
  • Page 12: Installing The Internal Manikin Adapter

    Internal Manikin Adapter sections with scissors, then place and test it for accurate detection of pad placement per local protocols. B. Installing the Internal Manikin Adapter M5088A with the Laerdal Resusci Anne HEARTSTART TRAINER M5085A...
  • Page 13: Installing Or Replacing The Training Pads Cartridge

    NOTE: Trainer will also operate with a modified Laerdal AED Anne. Contact Laerdal Medical for more information. Installing or Replacing the Training Pads Cartridge To install the training pads cartridge, follow these steps: 1. If a cartridge is currently installed in the Trainer, locate the latch at the top edge of the Trainer, and slide it to the side.
  • Page 14: Selecting A Training Scenario

    BATTERY when the Trainer is turned on, at the completion of a training scenario, and when the Trainer is turned off. The Trainer will continue to operate for a short time. Replace the batteries as soon as possible to ensure correct Trainer behavior. HEARTSTART TRAINER M5085A...
  • Page 15: Running A Trainer Scenario

    Running a Training Scenario To run the selected training scenario, treat the Trainer exactly as you would the HS1 Defibrillator. 1. Pull the handle or press the green On/Off button to turn on the Trainer. 2. Remove the clear protective lid from the training pads cartridge and set it aside.
  • Page 16: Standard Scenarios

    SYMBOL MEANING Shockable rhythm detected by defibrillator. Non-shockable rhythm detected by defibrillator. Pause provided by defibrillator. Pads problem detected by defibrillator. A Precaution is a condition, hazard, or unsafe practice that could result in minor personal injury. HEARTSTART TRAINER M5085A...
  • Page 17 NUMBER SCENARIO OVERVIEW SCENARIO DETAILS Shockable rhythm, one shock • Trainer detects a shockable needed for conversion rhythm, instructs user to deliver a † shock • Trainer detects a non-shockable ‡ rhythm Shockable rhythm, multiple shocks • Trainer detects a shockable needed for conversion rhythm, instructs user to deliver a †...
  • Page 18 Scenario will not progress unless a simulated shock is delivered. ‡ When the Trainer detects a non-shockable rhythm, it instructs the user to care for the patient and provide CPR if needed. The shock series used in the Trainer is always three shocks. HEARTSTART TRAINER M5085A...
  • Page 19: Trainer Performance During Use

    CPR and assessing the patient. Trainer Performance during Use The Philips HeartStart Trainer is designed for use in an automated external defibrillator training class taught by a qualified instructor. When a training scenario is running, the Trainer mimics the behavior of the HS1 Defibrillator during actual emergency use.
  • Page 20: Repacking The Training Cartridge

    (B). Then place the pads over the wound wires in the cartridge. The chest pad should be on top, and the tab ends of the pads at the curved end of the cartridge (C). Fold the tab down over the top pad (D). HEARTSTART TRAINER M5085A...
  • Page 21: Maximizing Training Pads Cartridge Service Life

    5. Press the resealable film cover over the cartridge, with the tab end at the top (E). Be sure to fold the tab back down under the cover (F). 6. Fit the hooks on the bottom of the clear hard protective lid into the holes at the bottom edge of the cartridge, and press the lid and handle down (G).
  • Page 22: Troubleshooting The Trainer

    Trainer. Trainer’s voice instructions One of the batteries is installed Be sure to install batteries backwards. according to the diagram in the high-pitched and too rapid. Trainer’s battery compartment. HEARTSTART TRAINER M5085A...
  • Page 23 Intentionally blank.
  • Page 24 P O W E R S A V E L I F E Philips Medical Systems is part Philips Medical Systems United States of Royal Philips Electronics Philips Medical Systems 2301 Fifth Avenue, Suite 200 Seattle, WA, USA 98121 (800) 263-3342...

Comments to this Manuals