Unit Reset; Table 5-4 Factory Default Values - Honeywell DirectLine DL422 User Manual

Sensor module for meredian ii orp electrodes
Table of Contents

Advertisement

5.3 Unit Reset

Overview
Unit Reset initializes all of the DirectLine® Sensor's calibration and configuration data to
factory default values.
Procedure
• From the Online ORP display, press and hold the
until the "rSEt" appears on the display (minimum of 10 seconds).
• "rSEt" will remain on the display until reset is complete. Next, the firmware version
number appears briefly and the unit then returns to the Online ORP display.
Zero Offset
Noise Suppression Frequency Selection
Output Configuration – 0 % Range Value
Output Configuration – 100 % Range Value
Output Configuration – 0 % Calibration
Output Configuration – 100 % Calibration
1/03

Table 5-4 Factory Default Values

Data
DirectLine® DL422 Sensor Module User Manual
and
buttons simultaneously
2
1
Default Values
0 mV
60 Hz
–1600 mV
1600 mV
4.00 mA typically
20.00 mA typically
Operation
19